EKO INSTRUMENTS CO., LTD. I-V Curve Checker MP-11 Instruction Manual Ver.7
Pg. 40
8 - 4 . S e t t i n g M e n u
Setting Menu screen appears when
“CONFIG” is selected from the Main Menu screen.
There are following options on the Setting Menu screen.
1. Range Setting
Manual setting is used when taking only the I-V curve measurement without using Sensor Unit and just taking
the measurement range manually.
Enter the measurement range of the subjected PV module or array for measurements.
From below contents, the ranges can be selected using the Left/Right buttons.
Voltage Range Setting
Voltage Range: 100V, 600V, 1000V, AUTO
(Default setting:
“AUTO”)
Current Range Setting
Current Range:
2A, 10A, 30A, AUTO
(Default setting:
“AUTO”)
2. Measurement Parameter Setting
Measurement Parameter Setting screen is displayed when the
“MEAS PAR” is selected from the Setting
Menu screen. There are following options for the Measurement Parameter Setting screen.
EDIT:
Edit parameter
SELECT:
Select registered parameter
ERASE:
Delete parameter
RANGE:
Range Setting
MEAS PAR:
Measurement Parameter Setting
AUTO MEASUREMENT:
Auto Measurement Setting
SYSTEM:
Environmental Condition Setting
Measurement Parameter Setting
MEAS PAR
ERASE
SELECT
EDIT
Range Setting
RANGE
Current Range AUTO
Voltage Range AUTO
Setting Menu