background image

21

Appendix

English

Technical Descriptions

Electromagnetic emissions

The RadiForce series is intended for use in the electromagnetic environment specified below.

The customer or the user of the RadiForce series should assure that it is used in such an environment.

Emission test

Compliance

Electromagnetic environment - Guidance

RF emissions 

CISPR11 / EN55011

Group 1

The RadiForce series uses RF energy only for its internal function.

Therefore, its RF emission are very low and are not likely to cause any interference in 

nearby electronic equipment.

RF emissions 

CISPR11 / EN55011

Class B

The RadiForce series is suitable for use in all establishments, including domestic 

establishments and those directly connected to the public low-voltage power supply 

network that supplies buildings used for domestic purposes.

Harmonic emissions

 

IEC / EN61000-3-2

Class D 

Voltage fluctuations / 

 

flicker emissions

 

IEC / EN61000-3-3

Complies

Electromagnetic immunity

The RadiForce series has been tested at the following compliance levels according to the testing requirements for 
professional healthcare facility environments defined in IEC / EN60601-1-2.

Customers and users of the RadiForce series must ensure that the RadiForce series is used in the following 
environments:

Immunity test

Test level for 

professional  

healthcare facility  

environments 

Compliance level

Electromagnetic environment - 

Guidance

Electrostatic 

discharge (ESD) 

IEC / EN61000-4-2

±8 kV contact discharge

±15 kV air discharge

±8 kV contact discharge

±15 kV air discharge

Floors should be wood, concrete or ceramic tile. 

If floors are covered with synthetic material, the 

relative humidity should be at least 30%.

Electrical fast 

transients / bursts 

IEC / EN61000-4-4

±2 kV power lines

±1 kV input / output lines

±2 kV power lines

±1 kV input / output lines

Mains power quality should be that of a typical 

commercial or hospital environment.

Surges 

IEC / EN61000-4-5

±1 kV line to line 

±2 kV line to ground

±1 kV line to line 

±2 kV line to ground

Mains power quality should be that of a typical 

commercial or hospital environment.

Voltage dips, short 

interruptions and 

voltage variations on 

power supply input 

lines  

IEC / EN61000-4-11

0 % U

(100 % dip in U

T

0.5 cycles and 1 cycle

70 % U

T

 (30 % dip in U

T

25 cycles

0 % U

T

 (100 % dip in U

T

)  

5 sec

0 % U

(100 % dip in U

T

0.5 cycles and 1 cycle

70 % U

T

 (30 % dip in U

T

25 cycles

0 % U

T

 (100 % dip in U

T

)  

5 sec

Mains power quality should be that of a typical 

commercial or hospital environment. If the user 

of the RadiForce series requires continued 

operation during power mains interruptions, it 

is recommended that the RadiForce series be 

powered from an uninterruptible power supply 

or a battery.

Power frequency 

magnetic fields

 

IEC / EN61000-4-8

30 A/m

(50 / 60 Hz)

30 A/m

Power frequency magnetic fields should be at 

levels characteristic of a typical location in a 

typical commercial or hospital environment. The 

product should be kept at least 15 cm away 

from the source of power frequency magnetic 

fields during use.

Содержание RadiForce MX194

Страница 1: ...urself with safe and effective usage Please retain this manual for future reference For monitor adjustment and settings refer to the Installation Manual For the latest product information including th...

Страница 2: ...unit This product has been adjusted specifically for use in the region to which it was originally shipped If operated outside this region the product may not perform as stated in the specifications N...

Страница 3: ...n of the Caution Statements Symbols on the unit Symbol This symbol indicates Power button Press to turn the monitor s power on or off Alternating current Alerting electrical hazard CAUTION Refer to SA...

Страница 4: ...not place in any form of transportation ships aircraft trains automobiles etc Do not place in dusty or humid environments Do not place in locations where water may be splashed on the screen bathrooms...

Страница 5: ...and install the unit securely Otherwise the unit may become detached resulting in injury and or equipment damage Before installation make sure that desks walls or any other installation surface has a...

Страница 6: ...tall the unit in a place with poor ventilation or inadequate space Do not use the unit laid down or upside down Blocking the ventilation slots prevents proper airflow and may result in fire electric s...

Страница 7: ...even after a short time period has elapsed depending on the displayed image If this occurs changing the image or leaving the power off for a few hours may solve the problem The backlight of the LCD p...

Страница 8: ...ng change Using the monitor with the control buttons locked is recommended after adjusting the screen of the monitor For details on how to set refer to the Installation Manual on the CD ROM Cleaning P...

Страница 9: ...and Functions 11 Chapter 2 Installation Connection 12 2 1 Before Installing the Product 12 Installation Requirements 12 2 2 Connecting Cables 13 2 3 Turning On the Power 14 2 4 Adjusting the Screen He...

Страница 10: ...l position for easy operation and minimizing fatigue Tilt 30 up 0 down swivel 35 right 35 left adjustable height 100 mm 1 2 Package Contents Check that all of the following items are included in the p...

Страница 11: ...e operation guide and the menus refer to the Installation Manual on the CD ROM 3 button Turns the power on or off 4 Power Indicator Indicates the monitor s operation status Green Operating Orange Powe...

Страница 12: ...this product on a lacquer coated desk the color may adhere to the bottom of the stand due to the composition of the rubber Check the desk surface before use Installation Requirements When installing t...

Страница 13: ...ns page 17 before connecting the PC 1 Connect signal cables Check the shapes of the connectors and connect the cables After connecting the DVI cable or the D Sub cable to the monitor tighten the faste...

Страница 14: ...ompletely when not using the monitor Note When turning on the monitor and PC for the first time with the analog signal the Auto Adjustment function works to adjust the clock phase and display position...

Страница 15: ...r reboot the PC 2 The message below appears This message appears when the signal is not input correctly even when the monitor functions properly This message appears when no signal is input Example Th...

Страница 16: ...400 69 Hz 71 Hz D Sub 50 0 Hz 75 0 Hz Synchronization Signal Separate Dot Clock Max DVI DisplayPort 108 MHz D Sub 135 MHz USB Port Upstream x 1 Standard USB Specification Revision 2 0 Power Input 100...

Страница 17: ...nnected and DP Power Save On 4 2 Compatible Resolutions The monitor supports the following resolutions For Digital Signal Input DisplayPort DVI Resolution Vertical Scan Frequency 640 x 480 60 Hz 720 x...

Страница 18: ...or EIZO FP 702 Arm Stand EIZO LS HM1 D Dual height adjustable stand EIZO LA 011 W Wall hung arm EIZO AAH 02B3W Wall hung arm Calibration Kit EIZO RadiCS UX2 Ver 4 6 3 or later Network QC Management So...

Страница 19: ...etic waves that could influence limit or result in malfunction of the monitor Install the equipment in a controlled environment where such effects are avoided Classification of Equipment Type of prote...

Страница 20: ...ent or stacked use is necessary the equipment or system should be observed to verify normal operation in the configuration in which it will be used When using a portable RF communication equipment kee...

Страница 21: ...idance Electrostatic discharge ESD IEC EN61000 4 2 8 kV contact discharge 15 kV air discharge 8 kV contact discharge 15 kV air discharge Floors should be wood concrete or ceramic tile If floors are co...

Страница 22: ...site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol Note 1 UT is the a c mains voltage...

Страница 23: ...aximum power W Minimum separation distance m IEC EN60601 test level V m Compliance level V m 385 380 390 TETRA 400 Pulse modulation b 18 Hz 1 8 0 3 27 27 450 430 470 GMRS 460 FRS 460 FM 5 kHz deviatio...

Страница 24: ...Hz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 7 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output po...

Страница 25: ...ce to radio communications However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television receptio...

Страница 26: ...ZO Corporation All rights reserved 1st Edition March 2018 Printed in Japan 00N0N072A1 IFU MX194 6 http www eizoglobal com 153 Shimokashiwano Hakusan Ishikawa 924 8566 Japan 8 5B Siemensallee 84 76187...

Отзывы: