Appendix
13.1 Information on electromagnetic compatibility (EMC)
CuratOR LX491W
Instructions for Use, 02/2018
61
Information and manufacturer’s declaration – electromagnetic radiation
The CuratOR LX491W is intended for use in the electromagnetic environments noted below.
The customer or user of the CuratOR LX491W has to ensure that the device is used in such an envi-
ronment.
Interference im-
munity test
IEC 60601
Measurement level
Compliance level
Information regarding the electro-
magnetic environment
Electrostatic dis-
charge (ESD)
IEC/EN
61000-4-2
±8 kV contact
±15 kV air
±9 kV contact
±15 kV air
It is recommended to use the device
on wood, concrete, or ceramic floors.
If the floor is made of synthetic mate-
rial, the relative humidity should be at
least 30%.
Fast transient
electric distur-
bances (burst)
IEC/EN
61000-4-4
±2 kV (5 kHz and
100 kHz burst fre-
quency)
for power lines
±1 kV (5 kHz and
100 kHz burst fre-
quency)
for input/output
lines
±3 kV (5 kHz and
100 kHz burst fre-
quency)
for power lines
±2 kV (5 kHz and
100 kHz burst fre-
quency)
for input/output lines
The power supply quality has to corre-
spond to that of typical industrial envi-
ronments or hospitals.
Surge voltage
IEC/EN
61000-4-5
±1 kV line(s)
against line(s)
±2 kV line(s)
against ground
±2 kV line(s)
against line(s)
±3 kV line(s)
against ground
The power supply quality has to corre-
spond to that of typical industrial envi-
ronments or hospitals.
Voltage dips,
brief interrup-
tions, and volt-
age fluctuations
along power
supply lines
IEC/EN
61000-4-11
<5 % V
T
(100 % dip
in V
T
) for 0.5 cycles
40 % V
T
(60 % dip
in V
T
) for 10 cycles
70 % V
T
(30 % dip
in V
T
) for 25 cycles
<5% V
T
(100 % dip
in V
T
) for 250 cycles
(5 seconds)
<5 % V
T
(100 % dip
in V
T
) for 0.5 cycles
40 % V
T
(60 % dip
in V
T
) for 10 cycles
70 % V
T
(30 % dip
in V
T
) for 25 cycles
<5% V
T
(100 % dip
in V
T
) for 250 cycles
(5 seconds)
The power supply quality has to corre-
spond to that of typical industrial envi-
ronments or hospitals.
If the monitor has to continue opera-
tion even if the power supply is inter-
rupted, it is recommended to connect
the device to an uninterruptible power
supply or battery.
Magnetic fields
with energy
technology fre-
quencies (50/60
Hz)
IEC/EN
61000-4-8
30 A/m
30 A/m
The magnetic fields with energy tech-
nology frequencies must be in an area
that is representative of a typical loca-
tion in a typical industrial environment
or hospitals.
Note:
V
T
is the alternating current voltage before application of the measurement level.
The measurement levels were established according to IEC60601-1-2 4th Edition - Professional