Technical specifications
10.5 Climatic conditions
CuratOR LX491W
Instructions for Use, 02/2018
57
10.5 Climatic conditions
In operation
Temperature range
+5 °C - +40 °C ambient temperature
Temperature gradient
Max. 5 °C/h, no condensation
Air pressure
700 - 1060 hPa (3000 m - -360 m)
For transport and storage (packed)
Temperature range
-20 °C - +60 °C ambient temperature
Temperature gradient
Max. 5 °C/h, no condensation
Humidity
10 - 90 %, non condensing, at 25 °C
Air pressure
200 - 1060 hPa (12000 m - -360 m)
10.6 Safety regulations
This device bears a CE mark in accordance with the provisions of Council Directive 93/42/
EEC and 2011/65/EU.
Safety standards
•
IEC 60601-1 (Edition 3.1)
•
IEC 60950-1:2005 (Second Edition) + A1:2009 +
A2:2013
•
CAN/CSA - C 22.2 60601-1:08 (Third Edition)
•
GB 4943.1 (non-tropical, altitude < 2000 m)
Protection class
Protection class I
Degree of protection according to
EN 60529
Front: IP65
Back: IP20
Medical product classification
Class I according to 93/42/EEC, Appendix IX