10 EMC Tables
Page 35 / 41
Guidance and manufacturer’s declaration – electromagnetic immunity
The
AirPex
is intended for use in the electromagnetic environment specified
below. The customer or the user of the
AirPex
should assure that it is used
in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment -
guidance
Conducted dis-
turbances
induced by RF
fields
IEC 61000-4-6
3 V
0.15 MHz
–
80 MHz, 6 V
in ISM bands
be-tween
0.15 MHz
and 80 MHz,
80 % AM at
1 kHz
3 V/m, 80
MHz
– 2,7
GHz, 80 %
AM at 1 kHz
3 V
Portable and mobile
RF communications
equipment should be
usedno closer to any
part of the
AirPex
,
including cables, than
the recommended
separation distance
calculated from the
equation applicable to
the frequency of the
transmitter.
Recommended
minimum separation
distances
See the RF wireless
communication
equipment table in
"Recommended
Radiated RF
EM fields
IEC 61000-4-3
3V/m