6 Operation
Page 17 / 41
6.3
Operation and not suitable condition
Press the back cover of the file clip to make
the hook of the file clip stick out. And hook the
metal handle of the root canal file. Release the
pressure and use the elasticity of the file clip
to complete the connection between the file
clip and the root canal file.
NOTE
When connecting the root canal file, make
sure that the file clip and the root canal file
handle are basically perpendicular, otherwise
the chuck of the file clip will be easily
damaged.
This equipment does not include the root canal
file. We need to buy another suitable model
according to the clinical needs. The metal part
of the root canal file should be well conductive.
When the file clip
can’t enter the patient’s
mouth, the file clip can be replaced by the
extension cord with the touch probe. Press the
touch probe on the metal handle of the root
canal file to complete the connection between
the touch probe and the root canal file.