26
9.2 Replace the battery
1. Dismantle battery
According to the direction of the arrow, toggle “Push button of battery storage”, removed “Upper cover of battery
storage” as shown in figure, take out the battery.
2. Install battery
Put the battery into battery storage of main unit; cover the “Upper cover of battery storage” into the corresponding
position, and then toggle the push button in the opposite direction, lock the “Upper cover of battery storage”.
Upper cover of battery storage
Battery
Push button of battery storage
Fig. Dismantle, install the battery
9.3 Troubleshooting
User shall prepare a proper maintenance plan featuring regular servicing and examination in order to ensure normal
instrument performance. If any abnormity occur, timely contact International Trade Dept of Eickemeyer for support.
If the following problems occur on starting up the machine, try to make corrections following the method in the table. If the
problem remains unsolved, contact International Trade Dept of Eickemeyer for support.
Trouble
Correction
Power-on light is not on and no screen
display is present when starting the machine.
1. Check power supply.
2. Check power cable and connections.
3. Check power adapter.
Character and grey scale are displayed, but
no ultrasonic image on the screen.
Probe is not properly connected. Turn off the power and reconnect the
probe.
Intermittent stripe, snow, or far-field
interference appears on screen.
1. Check power supply.( spark interference present)
2. Check environment.(source of interference around the machine,
such as electric motor, ultrasonic atomizer, automobile, computer or
other interference )
3. Check power plug/socket of the instrument or probe connectors.
They shall be properly contacted.
Image display is not clear.
1. Adjust the total gain, near field, far field.
2. Adjust the brightness and contrast level.
Control panel malfunction
Restart the system by turning off the main unit power.
9.4 System maintenance
To ensure the performance and safety of instrument, it must be checked after using 1 year. When check the instrument,
please consult the factory-authorized distributor of Eickemeyer Company for service, as they need to have professional
technology engineers.
Inspect
sorts
Cleaning
Electrical safety
Mechanical safety
Image recording
Inspect
items
System inside
Earth resistance
Earth leakage current
Inspect the control panel
Installed machine of
Images in each mode