19
Press
key unfreeze and quit playback status.
Note: If the images appear abnormal, that is without enough storage time and the images have not been stored full.
6.13 Text Input
Operation steps:
1. Freeze image;
2. Press
key, the cursor is located behind name;
3. Press
key again or right click mouse, at the same time characters input menu will be shown at the bottom of the
screen:
Caps 0 1 2 3 4 5 6 7 8 9 a b c d e f g h
Shift i j k l m n o p q r s t u v w x y z
Press direction keys or operate mouse to move cursor to point to Caps, and then press
key or left click mouse to
achieve
capital and small letter conversion; If the cursor point to Shift, press
key or left click mouse again to achieve
the conversion between the letter and punctuation;
4. Press direction keys to choose “numbers” or “characters” and press
key to confirm; Or operate mouse to click
“numbers” or “characters” to input;
5. After inputting name, press
key and then press direction key
to move cursor to “ID” and press
key again to
input according to Step 4, the number (ID) only be made up of numbers, letters or underlines;;
6. After inputting ID, press
key and press direction keys to move cursor to image area and press
key again to input
according to Step 4;
7. If need modify the content, continuously press
key twice to quit annotation status, press direction keys
to
select “Clear”, at last press
key to clear all noted marks and retype;
8. Press
key to quit.
6.14 Case report processing center V3.1 software (optional)
6.14.1 Function Introduction
1. Connect computer with USB interface of main unit, ultrasonic images transmit to PC and realize images mass memory.
2. Support five types case report template edit and print.
6.14.2 Use Instruction
1. Follow the "USB Installation Instruction V3.1" of CD to install the "Case Report Processing Center V3.1" software.
2. Insert USB cable into "USB interface" of main unit, and then switch on the power of main unit.
3. Connect the other end of USB cable to "USB Interface" of computer.
4. Open the "ImageUsb.exe" file in the Case Report Processing Center V3.1, that is double click the logo
, and
click
, then click "Start", the current displaying image transmitted to PC.
5. Click "Save" and then input the file name, so the image saved to PC.
6. Press the
button to exit the Case Report Processing Center V3.1.