Explanation: 

1. When reading images, it must choose the image code with “*”. The stored and read out images are all black and white 

images. 

2. In the status for saving or reading images, press

key, can delete the image with “*” mark in main unit or in the U 

disk. 

 
6.11 Single dump and mass dump 
6.11.1 Single dump 

 

From main unit to U disk 

1. Read out the image in the main unit, “Img” with highlight status is displayed on lower right corner of the screen; 

2. Press

key, a “Dump” prompt appears on lower right corner of the screen; 

    3. Press direction keys

to select the image saving code in the U disk, such as choose “005”; 

    4. Press

key, “Dump” prompt disappears, until the “Dump” re-appears, the current image can be dumped in the frame for 

coded “005”. The dumped image code is preceded with asterisk “*”; 

    5. Press

key to quit dumping status and press

key to return to real-time status. 

 

From U disk to main unit 

1. Read out the image in the U disk, “Img” with highlight status is displayed on lower right corner of the screen; 

2. Press

key, a “Dump” prompt appears on lower right corner of the screen; 

    3. Press direction keys

to select the image saving code in the “LOCAL”, such as choose “006”; 

    4. Press

key, “Dump” prompt disappears, until the “Dump” re-appears, the current image can be dumped in the frame for 

coded “006”. The dumped image code is preceded with asterisk “*”; 

    5. Press

key to quit dumping status and press

key to return to real-time status. 

6.11.2 Mass dump 

 

From main unit to U disk 

1. Freeze the image, continuously press cine loop key

until enter “PAUSE” status; 

2. Press

key, “LOCAL -> U Disk Mass Dump?” prompt display at the bottom of the screen; 

3.  Press

key,  “Dumping”  prompt  appears,  images  are  mass  dumped  under  the  “DUMP”  folder  in  the  U  disk,  and  the 

number behind “PAUSE” character is the number of folder name of mass dumping. After finishing mass dumping, return 
to “PAUSE” status; 

4. Press

key to quit mass dumping status and press

key to return to real-time status. 

 

From U disk to main unit 

1. Plug U disk; 

2. Freeze the image, continuously press cine loop key

until enter “PAUSE” status; 

3. Press

key twice, “U Disk -> LOCAL Mass Dump?” prompt display at the bottom of the screen; 

4. Press

key, “Dumping” prompt appears, the number behind “PAUSE” character is the number of folder name of mass 

dumping. After finishing mass dumping, return to “PAUSE” status; 

5. Press

key to quit mass dumping status and press

key to return to real-time status. 

Explanation: Mass dump, which is dump images from main unit to “DUMP” folder in the U disk; can also dump the images 

under the “DUMP” folder in the U disk to main unit.

 

 

6.12 Cine loop 

In  real-time  mode,  the  system  is  always  saving  the  scanned  image.  The  playback  images  are  for  a  period  time  images 

before freeze. 

Freeze the image, continuously press

key three times to enter the automatic playback status; Press

key to enter 

pause status when playing back; Press direction keys

or operate mouse to view images frame by frame in pause status; 

Continuously press

key three times again to return to automatic playback status. In the process of saving and playback, 

relevant data of image frames saved and played are shown on the lower right corner of the screen.   

Press

key to return to frozen status.   

Содержание MAGIC 500

Страница 1: ...r s Manual Information for Users Users shall carefully read through this manual and fully understand the text before operating the equipment This manual shall be placed easy of access for future reference otherwise it may cause instrument damage or person injury ...

Страница 2: ...responsibilities of implied marketability or merchantability applicable to the product for a particular use Eickemeyer shall be responsible for the safety reliability and proper performance of this equipment on condition that the assembly extension readjustment modification or service of the equipment herewith related is done by technical personnel authorized or approved by Eickemeyer Medizintechn...

Страница 3: ...uipment To obtain the right to return the goods contact local dealer of Eickemeyer Company indicating product serial number on the packing case or parts serial number on the part product model and reason to return the product through the dealer to return the products or parts Manufacturer s Information EICKEMEYER MedizintechnikfürTierärzteKG Eltastraße8 78532Tuttlingen Germany email info eickemeye...

Страница 4: ... 96 EC The device accessories and the packaging have to be disposed of waste correctly at the end of the usage Please follow Local Ordinances or Regulations for disposal F 3 Contents C h a p t e r O n e Te c h n i c a l S p e c i f i c a t i o n s 1 C h a p t e r T w o S y s t e m O u t l i n e a n d S t r u c t u r e 2 2 1 S t r u c t u r e c o m p o s i t i o n o f t h e i n s t r u m e n t 2 2 ...

Страница 5: ... s t m e n t 1 0 6 5 4 D y n a m i c R a n g e A d j u s t m e n t 1 0 6 6 Selection of Magnification Frequency Frame Correlation Image Post process Edge Enhancement Focus Number L o c a l Z o o m a n d D e p t h H e i g h t e n 1 0 6 6 1 M a g n i f i c a t i o n S e l e c t i o n 1 0 6 6 2 F r e q u e n c y S e l e c t i o n 1 0 6 6 3 F r a m e C o r r e l a t i o n S e l e c t i o n 1 1 6 6 4 I...

Страница 6: ...4 S y s t e m m a i n t e n a n c e 2 2 C h a p t e r Te n S t o r a g e a n d T r a n s p o r t a t i o n 2 3 C h a p t e r E l e v e n S a f e t y C l a s s i f i c a t i o n 2 3 A p p e n d i x S y s t e m B l o c k D i a g r a m 2 4 Chapter One Technical Specifications 1 1 Technical Data 1 Grey scale 256 2 Monitor 5 7 LCD 3 Main power supply 100 240V frequency 50 60Hz 4 Rating power 40VA 5 Mai...

Страница 7: ...tch map 2 3 Parts of the probe Take 6 5MHz animal transrectal linear array probe for example Fig Parts name of 6 5MHz animal transrectal linear array probe Name Function 1 Acoustic lens Toconvertelectricsignaltoultrasonicsignalbasedonprincipleofconversepiezoelectric effect Theultrasonicsignal afterenteringthehumanbody isreflectedasechowaveand convertedtoelectricsignalagain Theacousticlensisonthepr...

Страница 8: ...high frequency linear array probe 3 5 MHz convex array probe 2 Video printer P93W S 3 Mouse Chapter Four Operation Condition 4 1 Power supply Mains voltage 100 240 10 Mains frequency 50 60Hz 1Hz 4 2 Operation Environment Ambient temperature 10 40 Relative humidity 30 75 without condensation Atmospheric pressure 700hPa 1060hPa 4 3 Storage and Transport Ambient temperature 20 55 Relative humidity 30...

Страница 9: ...faults Warning 1 When instrument works abnormally do stop working turn off the power and check the reason then contacts the EICKEMEYER Company about it 2 Turn off power and pull out of the plug from socket after each ultrasonic diagnostic operation 3 It is forbidden to drag and press the power and probe cables emphatically regularly inspect whether there is pull apart and bareness if there is the ...

Страница 10: ...ng or using the probe make sure that the probe connecting cable and connector are in normal condition free of cracks or drop Use of defective probe may cause electric shock Warning 1 The Eickemeyer ultrasonic probe shall be connected to the dedicated Eickemeyer ultrasonic system only Select proper probe model according to the relevant instructions of ultrasonic diagnostic system 2 Check the ultras...

Страница 11: ...core power supply It connects with the protective earth line when power plug inserts into its socket 5 6 Video printer installation 1 Connect the equipotential terminal of the video printer to the earthing 2 Connect one end of the video connection line to the video printer and the other end to the video output interface on the left of main unit 3 Insert one end of power plug jack of the video prin...

Страница 12: ... machine 1 No abnormal voice strange smell and overheating appear 2 Check the machine to ensure a normal start up The power indication light is on and startup picture is shown on the screen The machine will be then automatically set in B mode 3 Check the acoustic lens for abnormal heat when the probe is in use This can be done by hand touching the probe to feel the temperature of the lens 4 Check ...

Страница 13: ... 192 224 and 255 5 Press key to quit setting interface 6 2 5 Hospital Name Setting 1 Press keytoentermain menuandpressdirectionkeys tomovecursorto Preset 2 Press direction keys to enter setting interface 3 Press direction keys to move symbol to point to Hospital 4 Press key the cursor is located above NAME at the same time characters input menu will be shown at the bottom of the screen Caps 0 1 2 ...

Страница 14: ...nd a single framed B mode image is displayed 6 4 2 B B Mode 1 In real time mode press key to enter B B mode 2 B B image switch Press key to enter main menu and press direction keys to move the cursor to B B Mode and then press direction keys to switch image display Or right click mouse to switch image display the selected image is activated and the other one is frozen 3 In real time mode press key...

Страница 15: ...tion Image Post process Edge Enhancement Focus Number Local Zoom and Depth Heighten 6 6 1 Magnification Selection In real time mode press key to enter main menu and move cursor to Zoom and then press direction keys to choose eight kinds of magnification 6 6 2 Frequency Selection In real time mode press key to enter main menu and press direction keys to move cursor to Freq in the display area Press...

Страница 16: ...on keys to realize image vertical reverse 6 7 3 Image Negative In real time mode press key to enter main menu and press direction keys to move cursor to Image POS and then press direction keys to realize negative function 6 7 4 Color Selection In real time mode press key to enter main menu and press direction keys to move cursor to COLOR in the display area and then press direction keys to realize...

Страница 17: ...lect the image code need to be saved in the U disk such as choose 004 5 Press key to appear Wait prompt After the prompt disappear the current image is saved in the frame for coded 004 The current image is saved under the specified folder in the U disk the folder name is patient number ID which is made up of underline letters or numbers without spaces file name is the selected saving code If user ...

Страница 18: ...s cine loop key until enter PAUSE status 2 Press key LOCAL U Disk Mass Dump prompt display at the bottom of the screen 3 Press key Dumping prompt appears images are mass dumped under the DUMP folder in the U disk and the number behind PAUSE character is the number of folder name of mass dumping After finishing mass dumping return to PAUSE status 4 Press key to quit mass dumping status and press ke...

Страница 19: ...nes 6 After inputting ID press key and press direction keys to move cursor to image area and press key again to input according to Step 4 7 If need modify the content continuously press key twice to quit annotation status press direction keys to select Clear at last press key to clear all noted marks and retype 8 Press key to quit 6 14 Case report processing center V3 1 software optional 6 14 1 Fu...

Страница 20: ...ace press key again the cursor will show or click left mouse the cursor will show 3 Press direction keys or operate mouse to move the mark to desired position press key or click left mouse to set the mark position as the start of the measurement or click right mouse to set the mark position as the start of the measurement 4 Press direction keys or operate mouse to move the mark to the end of the m...

Страница 21: ...surement and Calculation items Obstetric tables of the system including 1 BOVINE 2 EQUINE 3 OVINE 4 CANINE 5 FELINE 6 GOAT 7 LLAMA 8 SWINE which is reference for doctor 8 2 Measurement of Gestational Age GA and Estimated Date of Confinement EDC Follow the steps below 1 In B B B mode freeze the desired image press direction keys to choose OB in the display area 2 Press key to display the obstetric ...

Страница 22: ...ear all marks and data and quit the measurement status 8 3 Obstetric report In frozen status press key to enter main menu and press direction keys to move the cursor to the position Report Press direction keys to display obstetric report Press key to exit obstetric report status 8 4 Measurement items 1 Items measurable in B mode distance circumference area volume gestational age GA and estimated d...

Страница 23: ...n off the instrument and pull out the power supply wire before cleaning every instrument of the system It may cause electric shock if clean the system under power is on Warning There is no any water roof device in the system Do not splash any water or liquor into the system otherwise it will cause malfunction or electric shock Attention 1 Probe without cleaning disinfection or sterilization may be...

Страница 24: ...h dipped in moderate detergent and then air dry it 9 1 2 Use and maintenance for the charging battery 1 Only use charger and corresponding model of battery provided by Eickemeyer Company 2 The output port of adapter plugs into the input port of charger to charge the minimum charging time is 3 hours and the full charged battery can be used for 1 5 2 hours 3 The fore three times battery charging sho...

Страница 25: ...her with one off battery such as dry battery or different capability or different model or different brand battery which will lead to the leakage explosion or fire 13 Do not use the abnormal battery with particular smell or abnormal heat or distortion or turn colors or abnormal phenomena which will lead to the leakage explosion or fire 14 Do stop the charge and pull out the battery from the charge...

Страница 26: ... the machine 1 Check power supply 2 Check power cable and connections 3 Check power adapter Character and grey scale are displayed but no ultrasonic image on the screen Probe is not properly connected Turn off the power and reconnect the probe Intermittent stripe snow or far field interference appears on screen 1 Check power supply spark interference present 2 Check environment source of interfere...

Страница 27: ...ell ventilated and free of direct sunlight and corrosive gas 3 Shockproof measures have been taken inside the packing case to allow for transport by air railway land and sea The goods shall not be exposed to poor weather conditions like rain and snow nor shall the goods be placed upside down bumped knocked or over stacked Chapter Eleven Safety Classification 1 Classified according to electric shoc...

Страница 28: ...28 Appendix System Block Diagram Probe Transmission Reception Keyboard Microcomputer control system Clock Line field sync signal Collating DSC Process Signal synthesis Display ...

Страница 29: ...nik für Tierärzte KG Eltastraße 8 78532 Tuttlingen Germany tel 49 0 7461 96580 0 fax 49 0 7461 96580 91 mail export eickemeyer de web www eickemeyer de Information contained in this manual is subject to change without further notice ...

Отзывы: