5.8.2 Inspection After Start-up
Check the following items after starting the machine:
1. No abnormal sound, strange smell and overheating appear.
2. Check the machine to ensure a normal start-up: The power indication light is on and the startup picture is shown on the
screen. Then the machine will be automatically set in B mode.
3. Check the acoustic lens for abnormal heat when the probe is in use. This can be done by hand touching the probe to feel
the temperature of the lens.
4. Check the image to ensure trouble-free display (no excessive noise or fl icker).
5. Check the control panel to ensure normal operation condition.
6. Check the instrument to ensure that the phenomenon of local high temperature will not appear.
Attention!
• If the overheat acoustic lens is placed on the patient’s skin, heat injury may occur.
• Thoroughly clean the coupling gel on the probe surface each time after ultrasonic operation, or the coupling
gel may become hardened on the acoustic lens of the probe, deteriorating image quality.
5.9 System Reset
In case of abnormal screen display or no-working for system operation, try to restart the system by turning on / off the main
unit power.