It is advisable to recharge the battery at least once
a week in order to not compromise charge
capacity.
A reset button (6), which can be operated via a
fine tip, restores the standard conditions of the
device.
4.5.
Wall mounting
Stet Cube can be attached to a vertical wall with the
wall bracket provided (11). For wall mounting, it is
first of all necessary to fix the bracket to the wall at
the correct height, using the screws provided.
Stet Cube can then be fitted to the bracket via the
housings underneath the base (12). The system can
be removed from the bracket at any time by sliding
Stet Cube upwards. If fitted near to an electricity
outlet, Stet Cube can also be recharged directly
from its position on the wall. “Vertical” disinfection
operations are described in figure D (phases I
→
II
→
III). It is recommended to always place the head
of the stethoscope within the transparent housings
(2) (4) to ensure it is held firmly in place even with
the lid open.
5.
OPERATING RANGES
Stet Cube ensures effective disinfection throughout
the shelf life of the UV-C LED (more than 50.000
uses).
During hygienization phases, the system defines
automatically a T1 step (standard), lasting 3
minutes + T2 step, adding an extra 2 minutes to
disinfect stethoscope more deeply.
6.
LIGHT SIGNALS
The operating lights indicate:
-Blue light = UV-C LED operation
-YELLOW light = battery operation
6.1 Summary table of light signals
7.
LIMITATIONS
Stet Cube can be used with almost all stethoscopes.
The incomplete closure of the lid may, however,
prevent the device from working
8.
IMPORTANT WARNINGS
We recommend you follow the instructions
below.
Keep away from heat, sunlight, moisture and
water.
It is recommended to dry the membrane of
the stethoscope before inserting it into Stet Cube,
to avoid the formation of condensate and ensure
optimal disinfection.
Keep the Stet Cube away from sources of
electromagnetic waves and ionising radiation (MRI,
x-ray, microwave, etc.).
Do not tamper with or force the closure
mechanism, to avoid activating the UV-C LED with
the lid open and/or the stethoscope placed
incorrectly.
Should the UV-C LED be activated when the lid
is open, never direct the UV-C LED towards skin and
eyes.