GES 400 MR Series User Manual | 8100401-55
•
2015-07-22
49
4.5
GES External Power Supply
CAUTION:
•
The medical grade GES External Power Supply is
MR UNSAFE
and
not to be connected or used within the MR
environment
. While inside the MR imaging room, power the
FICS by its internal battery.
•
DO NOT unplug the GES External Power Supply from the FICS
while the internal battery is being recharged.
NOT TO BE USED INSIDE THE MR ENVIRONMENT.
Used only
outside of the MR
environment
to:
•
recharge the FICS’s
internal battery
•
power the amplifier for
EEG-only acquisition
4.6
Isolation Transformer
All EGI systems include one or more isolation transformers (medical
grade Toroid W Series), and all EGI components,
except the FICS
,
must plug into an isolation transformer to provide the attached
components with isolation from mains power for patient safety.
Note:
If more than one isolation transformer is included, it is used to isolate
any optional event related potential (ERP) components (for research only)
from the EEG components.