Immunity Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic
Environment—
Guidance
Voltage dips,
short
interruptions,
and voltage
variations on
power-supply
input lines
IEC 61000-4-11
< 5% U
T
(> 95%
dip in U
T
) for
0.5 cycle
40% U
T
(60% dip
in U
T
) for
5 cycles
70% U
T
(30% dip
in U
T
) for
25 cycles
< 5% U
T
(> 95%
dip in U
T
) for
5 sec.
< 5% U
T
(> 95%
dip in U
T
) for
0.5 cycle
40% U
T
(60% dip
in U
T
) for
5 cycles
70% U
T
(30% dip
in U
T
) for
25 cycles
< 5% U
T
(> 95%
dip in U
T
) for
5 sec.
Mains power
quality should be
that of a typical
commercial or
hospital
environment. If the
user of the GES
requires continued
operation during
power mains
interruptions, it is
recommended that
the GES be
powered from an
uninterruptible
power supply or a
battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital
environment.
Note:
U
T
is the AC mains voltage before application of the test level.
Содержание GES 400 MR
Страница 4: ...4 GES 400 MR Series User Manual 8100401 55 2015 07 22...
Страница 12: ...12 GES 400 MR Series User Manual 8100401 55 2015 07 22...
Страница 52: ...52 GES 400 MR Series User Manual 8100401 55 2015 07 22...
Страница 54: ...54 GES 400 MR Series User Manual 8100401 55 2015 07 22 5 1 EEG fMRI Figure 5 1 Typical EEG fMRI configuration...
Страница 56: ...56 GES 400 MR Series User Manual 8100401 55 2015 07 22...
Страница 64: ...64 GES 400 MR Series User Manual 8100401 55 2015 07 22...
Страница 65: ...GES 400 MR Series User Manual 8100401 55 2015 07 22 65...