1.4
Safety Warnings
All EGI MR conditional equipment has been tested for MR safety based on the
guidelines from ASTM F2503-05 (Standard Practice for Marking Medical Devices
and Other Items for Safety in the Magnetic Resonance Environment), which is
one of the international standards of the American Society for Testing and
Materials (ASTM). Based on these tests (including displacement force and
torque, as well as RF heating), EGI’s MR conditional equipment is considered
safe in the specified MR environment.
The HC GSN MR Net is the only MR conditional device that will enter the bore of
a magnet. All related equipment is labeled as either MR conditional or MR
unsafe to ensure that only MR conditional devices will be allowed into the
specified MR environment. See the following tested use conditions.