The soft-start mode can be invoked following a power outage or disconnection of power
from the pump.
Table 12
LED indicators
Refer to
Figure: Interface control panel
for LED locations on the pump dashboard.
Description
Function
Power indicator
Indicates that electrical mains supply to the pump is ON.
Run indicator
Indicates that the pump is running.
▪ LED continuously ON
▪ Pump runs at full speed
▪ LED flashing
▪ When changing speed
▪ LED OFF
▪ Pump is not running
Standby mode indicator
Indicates that the standby mode has been selected.
Standby speed increase in-
dicator
The indicator will blink with every short push of the
standby speed increase button. The indicator will remain
ON when maximum standby speed has been reached.
Standby speed decrease
indicator
The indicator will blink with every short push of the
standby speed decrease button. The indicator will remain
ON when minimum standby speed has been reached.
Auto-run indicator
Indicates that the auto-run mode has been selected.
Standby indicator LED
Once the standby function is selected the run indicator
will flash at 50% duty while the rotational speed is
changed. Once steady state running speed has been
reached both the run speed and standby indicators will
be illuminated.
Service indicator
Indicates that a service interval has been reached.
Alarm indicator
Indicates an alarm has been triggered.
6.3 Logic interface data
The pumps have a 15-way D-type logic interface connector located on the user interface
, (6)). The logic interface connector can be plugged directly
into the 200 W Turbo Instrument Controller (TIC) with an nXRi TIC cable and in
conjunction with controller extension cables.
For Turbo controller, or Turbo and Active Gauge controller (TAG), a suitable connector
mating half must be used (not supplied) to connect the nXRi pump to the customer
control system. Refer to and
for the electrical connections.
Table 13
Interface technical data
Connector*
15-way D-type (male)
Start, serial enable and remote enable:
▪ Enable control voltage: low (closed)
▪ Disable control voltage: high (open)
▪ 0 to 0.8 V d.c. (l
OUT
= 0.55 mA
nominal)
▪ 4 to 26.4 V d.c. (internal pull-up to a
Thevenin equivalent circuit: 5.3 V and
11 kOhms nominal)
05/2021 - ©Edwards Limited
Page 29
A90301880_C
A90301880_C - Operation