Operation
B800-00-880 Issue D
In parallel control mode, the pump will not accept serial stop commands but will accept all other serial commands.
When serial enable is active, the pump will run at standby speed if it has been commanded to do so by a serial standby
command. When Serial Enable is inactive and the RS485/RS232 switch is in RS232 position, the pump will run at
standby speed when the parallel standby line is active.
In serial control mode, the state of the parallel Start line will be ignored but the serial enable line provides an
interlock function as shown in
Figure 19
. This interlock function only operates with serial start commands and
therefore only in serial control mode. In serial control mode, the pump cannot be commanded to standby speed by
the standby line, instead a serial standby command must be used.
The parallel normal and analogue output signals provide valid pump status information at all times under mixed
parallel and serial operation. The analogue output voltages are described in
Table 10
.
Table 23 - Serial enable matrix
Serial enable
Active
Inactive
R
S48
5/
RS23
2 S
w
it
ch
RS232
Standby and fail lines are used for RS232 data. Standby and fail lines are used for parallel
Standby control and fail indication.
Parallel standby control and fail indication are
disabled.
Serial communications are disabled.
RS485
Standby and fail lines are used for RS485 data. Standby and fail lines are in (RS485) high
impedance state.
*
*
Pump target speed will switch between run speed and standby speed in response to any data driven onto
the RS485 bus by other devices connected to the bus. Activation of the parallel start line not recommended
with the RS485/232 switch in RS485 position and serial enable Inactive. Deactivation of the serial enable
line is not recommended with the pump running following a start from the parallel start line when the
RS485/232 switch is in the RS485 position.
Parallel standby control and fail indication are
disabled.
Serial communications are disabled.