© Edwards Limited 2007. All rights reserved.
Page 9
Edwards and the Edwards logo are trademarks of Edwards Limited.
Technical data
B722-01-880 Issue K
Table 4 - EXT250 performance data
DN100CF
DN100ISO-K
Pumping speed
*
Nitrogen
240 l s
-1
240 l s
-1
Helium
250 l s
-1
250 l s
-1
Hydrogen
190 l s
-1
190 l s
-1
Compression ration
Nitrogen
> 1 x 10
8
> 1 x 10
8
Helium
2 x 10
4
2 x 10
4
Hydrogen
1500
1500
Ultimate pressure
†
< 5 x 10
-10
mbar
< 5 x 10
-9
mbar
< 5 x 10
-8
Pa
< 5 x 10
-7
Pa
Maximum continuous inlet pressure
‡
Water-cooling at 15
o
C
with EXC100/120
1 x 10
-1
mbar
1 x 10
1
Pa
1 x 10
-1
mbar
1 x 10
1
Pa
with EXC300
3 x 10
-1
mbar
3 x 10
1
Pa
3 x 10
-1
mbar
3 x 10
1
Pa
Air-cooling at 35
o
C
3 x 10
-2
mbar
3 x 10
0
Pa
3 x 10
-2
mbar
3 x 10
0
Pa
Nominal rotational speed
60000 r min
-1
60000 r min
-1
Standby rotational speed
42000 r min
-1
42000 r min
-1
Starting time to 90% speed
with EXC120/120E
100 sec
100 sec
with EXC300
90 sec
90 sec
Recommended Controller
EXC100/120
EXC100/120
EXC120/E maximum input
250 VA
250 VA
EXC120/E normal power
60 VA
60 VA
Other compatible Controller
EXC300
EXC300
EXC300 maximum input
480 VA
480 VA
EXC300 normal power
60 VA
60 VA
Quiescent power consumption
25 W
25 W
*
Pumping speeds are without inlet-screen. Inlet-screens are supplied fitted and reduce speed by approximately
10%.
†
Ultimate pressure 48 hours after bakeout with 2-stage rotary vane backing-pump.
‡
Above this pressure, rotational speed drops below nominal.
Содержание EXT250 Series