Installation
A344-02-880 Issue J
3.9
Connect the pump outlet to your system
Connect the pump outlet-port to your outlet accessories or your exhaust treatment plant using the NW25 fittings
(supplied).
Take note of the following information before you connect to the pump outlet. Refer to
for details of the
accessories mentioned below.
z
The exhaust system must be configured so that the maximum pressure at the pump outlet does not exceed
0.5 bar gauge (1.5 bar absolute, 1.5 x 10
5
Pa) at full pump throughput.
z
In the following circumstances, we recommend that you fit an oil mist filter to the pump outlet:
z
if you use the pump with the gas-ballast control open
z
if you operate the pump with an inlet pressure greater than 10 mbar for extended periods
or
z
if you frequently pump down from atmospheric pressure.
The mist filter will trap the oil exhausted from the pump: you can re-use the oil if it is not contaminated.
3.10
Gas-ballast inlet connection
As shown in
(item 3), the pump has two gas-ballast inlets. We supply the pump with a blanking plug fitted
to one inlet: the other inlet is open. You can use either of the inlets, but you must fit the blanking plug to the inlet
you do not use.
You can leave the gas-ballast inlet open to the atmosphere. Alternatively, you can fit a gas-ballast control valve,
(available as an accessory, refer to
), for remote control of the gas-ballast supply. In either case, adjust the
gas-ballast control (
, item 4), to regulate the flow rate of the gas-ballast supply.
For more information about the use of gas-ballast, refer to Sections
and
3.11
Leak-test the system
Leak-test the system and seal any leaks found after you have installed the pump, to prevent leakage of substances
out of the system and leakage of air into the system.
WARNING
Connect the pump outlet to a suitable treatment plant to prevent the discharge of dangerous gases
and vapours to the surrounding atmosphere. Use a catchpot to prevent the drainage of
contaminated condensate back into the pump.