W601100880_A - Installation
Use differential measurement input (signal common and supply common wired
separately).
Potential difference between the supply common and the housing
16 V (over-
voltage protection).
FCC68, 8-pin connector
1. If a sensor cable is not available, one can be made. Refer to
Connect the sensor cable.
Figure 4
FCC69 8-pin sensor cable
1
Supply
2
Supply common ground
3
Signal output (measurement signal)
4
Gauge identification or remote zero adjust
5
Signal common
6
N.C.
7
N.C.
8
N.C.
case Connector case