Page 9
W601100880_A - Technical data
Technical data
Table 1
General
Measurement range
Accuracy
*
*
Non linearity, hysteresis, repeatability in the calibrated range at 25 °C ambient
operating temperature without temperature effects after 2 hours of operation.
W601xxxxx
1.0% of reading
Temperature effect on zero
0.02% F.S./°C
Temperature effect on span
0.02% of reading/°C
Resolution
0.05% F.S.
Gas type dependence
None
Mass
g
Table 2
Output signal
Output signal analogue (measurement signal)
Voltage range
-0.2 to +10.24 V
Measurement range
0 to +10 V
Relationship voltage-pressure
Linear
Output impedance
10
(short circuit proof)
Loaded impedance
> 10 k
Response time
100 ms
Table 3
Gauge identification
Gauge identification
Resistance 18 k
referenced to supply
common
WARNING: HIGH VOLTAGE
Risk of electrical shock. The gauge can only be connected to power supplies, instruments
or control devices that comply with the requirements of a grounded Protective Extra-Low
Voltage (PELV) and Limited Power Source (LPS), Class 2. The connection to the gauge has
to be fused.
*
*
Edwards controllers meet this requirement.
Table 4
Electrical
Supply voltage
at the gauge
+13 to +32 V d.c.
Class 2 / LPS
ripple
V
pp
Current consumption
< 20 mA
Power consumption
0.3 W
Fuse required
*
1 AT (slow), automatic reset (Polyfuse)
Содержание Barocel 7025M
Страница 1: ...W601100880_A Original Instructions Barocel 7025M Gauge INSTRUCTION MANUAL edwardsvacuum com ...
Страница 6: ...This page has been intentionally left blank Page 4 W601100880_A Contents ...
Страница 13: ...Page 11 W601100880_A Technical data Dimensions Figure 1 Dimensions mm ...
Страница 25: ...This page has been intentionally left blank ...
Страница 26: ...This page has been intentionally left blank ...
Страница 27: ......
Страница 28: ...edwardsvacuum com ...