W601100880_A - Installation
1. Position the gauge to prevent vibrations.
Note:
The gauge can be installed in any orientation. We recommend you install the gauge in a
horizontal to upright position and use a seal with a centring ring and filter to prevent the influx
of condensates and particles in the measurement chamber.
2. Install the gauge so that the buttons can be accessed with a pin for future
adjustments.
3. Remove the protective lid and connect the gauge to the vacuum system.
4. Keep the protective lid.