background image

TEMPERATURE CALIBRATION

AIR MOVEMENT WILL AFFECT THE TEMPERATURE READING.
WORK IN AN AREA WHERE THIS IS MINIMUM.
1. Use the 

DS299S 

to remove the temperature control knob.  Turn Temp. Control 

(A)

 so arrow

points to cord.

2. Use the 

SRW01

 to pull out heater assembly, PCB & power cord from handle.  Leave

enough power cord slack to install 

SRF30

.

3. Place heater assembly, PCB & power cord inside 

SRF30

.  Install tip that is clean and well

tinned.

4. With the use of the 

MS412 Temperature Caliration System

.  Follow set-up shown below.

5. Plug in tool

 

and by using the Pot Adjustment Tool supplied, turn Temp. Control 

(A)

 to

"solder-melt" temperature.

6. Apply a small amount of solder on the Tip, just enough to form a solder bead on top of

the Tip.  Place the center of the thermocouple wire of the on top of the solder bead.

7. Again, apply a small amount of solder on the center of the cross wire, just enough to

embed the center.

8. Turn Temp. Control 

(A)

 fully clockwise. Adjust Calibration Pot 

(B)

 until reading stabilizes at

800°F.

9. Assemble unit in original handle.
10. Before installing the Temp. Control Knob, peel-off pointer label from the Knob.
11. Using same MS412 set-up, install and adjust Temp. Control Knob so Meter will read desired

temperature.

12. Attach new pointer label in accordance with the reading of the Meter.

Calibration Pot 

(B)

SRF30

original handle

Temp. Control 

(A)

Tools Needed

•SRW01
•DS299S
•SRF30
•MS412
•Pot Adjustment Tool

SRW01

DS299S

MS412 Temperature
Calibration System

SDS100

SRF30

FX635

TI680 Meter

Pot Adjustment Tool

POINTER LABEL

Отзывы: