background image

 
 

 
 
 

 
 
 
 

 
 
 

12.1.2.2 – Once the selected password has been entered, touch the DONE key 
at the lower right hand location on the screen and a popup message will ask if 
the new password should be enabled. Touch YES to proceed or NO to cancel 

selection. Touch the DONE key to exit to the main menu. 
 
 

 
 
 
 

 
 
 

 
12.1.3 – Enabling the Password – Touch the Setup key and the Setup Access 
window will appear showing the Unlocked software key. To enable the 

password, touch the Unlocked key and a Lock Setup Access popup window 
question appears. Touch YES key to enable the password or NO key to exit. 
 

 
 
 
 

 
 
 

 
12.1.3.1 – If YES, the UNLOCKED key will now appear as Locked in a red box. 
Touch the EXIT key to return to the main menu. This action will require the 

password to be entered until unlocked or cancelled. 
 
 

 
 
 

 
 
 

Utilities – continued 

 

Содержание Accu-Touch

Страница 1: ...EDMUNDS GAGES Accu TouchTM Amplifier Instruction Manual...

Страница 2: ...h any air lines from the tooling to the front of the Accu touchTM docking unit if applicable noting their proper connection Step 3 Turn on the power switch located at rear of the amplifier near the po...

Страница 3: ...he unit comes with a one year warranty from date of factory shipment Damage caused by shipping handling unclean electrical power contaminated compressed air supply or mis use voids any warranty Tamper...

Страница 4: ...tion section 5 Setting pneumatic mag section 6 Software calibration section 7 Auto air shutoff setup section 8 Measuring section 9 Data offload setup section 10 Setup Report section 11 Utilities secti...

Страница 5: ...hose tooling connection Docking unit air gaging optional Rear view Power on off switch RS232C serial port LVDT ports rear USB port data offload or optional data send TIR reset footswitch LVDT jumper...

Страница 6: ...electrical power should be clean and stable without large fluctuations otherwise internal damage of electrical components can occur voiding warranty In such cases a isolation transformer is recommend...

Страница 7: ...Setup Access menu appears 1 2 1 Touch Measurement Setup setup now appears if unlocked or if password protection has not been enabled 1 2 2 The 11 selectable items related to for your use are availabl...

Страница 8: ...touch and select by touching one of four available methods of displaying the reading Live TIR PEAK PEAK 2 5 Description touch and enter letters numbers or characters using software keyboard to give a...

Страница 9: ...1 size Master 2 size touch and enter master 1 always maximum size master 2 always minimum size of the masters using software keyboard Use sizes from certifications if masters are out of original toler...

Страница 10: ...if not touch the Measurement Setup key to display the first check 3 1 2 Touch the Measurement software key to display the available Measurement formulas If one of the formulas shown can be used for th...

Страница 11: ...display System Setup 3 2 2 If the optional Formula Generator base version E7204A has been purchased the Formula Generator key will be present Touch this key to view the existing formulas 3 2 3 Existin...

Страница 12: ...la will be dependent upon the specific gaging application Gaging probes used within any system are strategically mounted to contact the work piece at specified locations to perform measurements The nu...

Страница 13: ...nd 1 000 for inside measurements The formulas would be A and A Checks can also be combine algebraically As an example air plug that measures a part ID at two elevations with one labeled as the large d...

Страница 14: ...n box Select minus under the Operators selection box Select C2 under the Check selection box 3 2 6 2 Once the formula has been entered touch the Verify key This will verify the formula has been entere...

Страница 15: ...on setups can be Exported Imported to USB drive 4 1 1 storing new setups touch Setup Setup Access menu appears Touch System Setup key System Setup menu now appears1 if unlocked or if password protecti...

Страница 16: ...own or up to the setup name required Once the required setup name is highlighted touch the Load from DB key and the system will respond with a warning before overwriting the existing setup If loading...

Страница 17: ...e setup name required Once the required setup name is highlighted touch the Export to USB key and the system will respond with a YES NO prompt to confirm If the setup already exists on the USB drive a...

Страница 18: ...setup name shown or scrolling down or up to the setup name required The system will respond with a YES NO prompt to confirm If the setup already exists in the internal database a warning before overwr...

Страница 19: ...Uncalibrated reading is now displayed in the digital window Place the master on the gage and read the value If the reading is within 200x of the resolution selected of the zero master calibrated size...

Страница 20: ...issue with the gage that needs to be addressed before continuing 5 1 3 1 Air gaging two master touch Calibrate key and Measure 1 will appear as previously shown 5 1 3 2 Touch Adv Cal Input Setup key a...

Страница 21: ...e air gage covering the nozzles If the digital reading is close or equal to the min master size proceed to software calibration Repeat steps 5 2 1 5 3 1 for other inputs If not proceed to step 5 3 2 M...

Страница 22: ...r an ID hole measurement airplug the reading will need to be smaller than the min master size For an O D shaft measurement air ring air snap the reading will need to be even larger than the reading fi...

Страница 23: ...master or single master types Combinations must be calibrated separately 7 2 1 Electronic gaging single master touch the Calibrate key and Measure 1 will appear 7 2 2 Place the master on the gage and...

Страница 24: ...or filter L bracket tubular nut 8 1 1 1 If already installed proceed to section 7 1 2 to setup the software 8 1 1 2 If the hardware has not been installed With the power and air pressure turned off di...

Страница 25: ...o toggle choices 8 1 2 3 Air Off Delay touch this key to select a time delay of between 0 to 90 seconds This time period sets the delay between the air supply will being reduced from the point that th...

Страница 26: ...ted in section 7 1 2 exit out to the main menu by touching the Done key in the lower right hand corner of the screen Touch it once to exit the Air Saver setup and touch it again to exit the System Set...

Страница 27: ...elect 9 1 2 Applications using live modes enable the radial bargraph and digital numbers to change in real time T I R PEAK or PEAK lock the radial bargraph digital value to that dynamic feature seen a...

Страница 28: ...beled Preference is listed below the Method key allowing Single Viewed or ALL to be selected If multiple measurements are setup all measurements can be offloaded at one time regardless of how many are...

Страница 29: ...ies screen The default file name is AccuTouch txt Basic format 7 features per check date time check name upper part limit lower part limit actual reading condition CR LF Examples One Check setup syste...

Страница 30: ...null modem 9 pin female x female cable is required to tie this amplifier to the mating device 10 3 3 When finished touch the Exit key to close the settings window and touch the DONE software key to ex...

Страница 31: ...following pins on the connector are utilized for communications Accutouch Null modem Cable External Device Pin 1 Chassis Ground Chassis Ground Pin 2 Receive RXD TXD Transmit Pin 3 Transmit TXD RXD Rec...

Страница 32: ...art of the screen The Setup Database Utilities screen will appear with a Offload Setup Report software key present at the lower left part of the screen 11 3 Touch the Offload Setup Report key and a po...

Страница 33: ...lled or recognized an error message will display indicating such Touch the OK key and exit out by touching the Done key twice until a compatible drive is available 10 1 6 If the drive is compatible an...

Страница 34: ...K 1 Sensor Type Electric Resolution 5 Decimal Units Inch Description CHECK 1 Mode LIVE Measurement INPUT A Nominal 1 574000 Master 1 Size 1 575400 Master 2 Size 0 000000 USL 1 575350 LSL 1 574950 Mast...

Страница 35: ...port without entering the password All other features are protected from change and or view 12 1 1 Touch the Setup software key and Setup Access menu appears 12 1 2 Entering a password As the password...

Страница 36: ...abling the Password Touch the Setup key and the Setup Access window will appear showing the Unlocked software key To enable the password touch the Unlocked key and a Lock Setup Access popup window que...

Страница 37: ...one key and a popup warning message will appear 12 1 4 1 Touch the YES key to remove the password protection option Touch the NO key to cancel and retain the option of enabling it 12 1 5 Removing pass...

Страница 38: ...re key and Setup Access menu appears 12 2 2 Touch the System Setup key System Setup menu displays with or without a password enabled 12 2 3 Touch the ABOUT key located at the upper right hand corner o...

Страница 39: ...wise if not locked touch the System Setup key and the System Setup menu displays 12 3 3 Touch the Set Date Time key located and a popup window will appear showing the Set Date Time feature 12 3 4 Touc...

Страница 40: ...ppears 12 4 2 Enter the password if required after touching the Locked key and then the DONE key Otherwise if not locked touch the System Setup key and the System Setup menu displays 12 4 3 Touch the...

Страница 41: ...s to be utilized or ignore allowing the software to fully load Once touched the Backup Restore menu will display 12 5 2 Backing up the last time the software was backed up will be listed near the Back...

Страница 42: ...overwrite of the running copy Touch the No key to stop the process and return to the Backup Restore menu 12 5 3 1 Touch the Yes key to continue When complete a Copy Complete message will appear Touch...

Страница 43: ...one type would be opposite of expected 5 Nominal setup available only on latest version software Nominal can be the zero between the max min master part sizes the zero of a single master application...

Страница 44: ...et to Multi Check When the type is set to Multi Check selections C1 C2 C3 C4 will be available to enter into the formula For example if Check 1 and Check 2 measure a shaft diameter at two different po...

Страница 45: ...9 9 Max function be used in Multi Check formula max C1 C2 will return greater of C1 C2 min Minimum value example max 4 9 4 Min function be used in Multi Check formula min C1 C2 will return lesser of C...

Страница 46: ...r user confirmation Verify Formula The formula must be verified before the screen can be closed and the formula saved permanently Whenever the formula has been changed the Verify button will be highli...

Страница 47: ...LIVE PEAK PEAK TIR The MAG setting on the A E signal conditioning module sets the gain between the max min sizes A Gage is a mechanical device used to measure part characteristics Gage Readings are t...

Страница 48: ...ment variation of a gage when used by one operator or under one set of environmental conditions Reproducibility is the variation in measurement averages of a gage when used by more than one operator o...

Страница 49: ...Improper connection to amplifier Faulty LVDT Input setup incorrect A E input not responding Incorrect or no air supply pressure Faulty regulator Improper connection to docking unit Input setup incorre...

Страница 50: ...Edmunds Gages 45 Spring Lane Farmington CT 06032 USA tel 860 677 2813 fax 860 6774243 info edmundsgages com WWW EDMUNDSGAGES COM Revision 0 4 5 11 Revision 2 4 1 15...

Отзывы: