4-21
LVDT Module Installation
1) Turn off the main power switch on the rear of the Trendsetter II and unplug the power
cord.
Power to the column must be turned off prior to installing or removing a signal
conditioning module.
2) Ensure all jumpers and switches are properly set for the application to be run.
3) Slide the LVDT module into the lower bay and secure it in position with the two
thumb screws on the front panel.
Note:
Ensure the card is fully inserted into the
lower bay and that the connector on the rear of module is properly seated in the
mating plug in the Trendsetter II.
4) Connect one or two LVDTs to the A and/or B plugs on the front of the module.
5) Plug in the Trendsetter II power cord.
6) Turn on the power switch on the rear of the Trendsetter II.
7) If necessary reprogram the scale, range, mode, etc. for the new application. See the
"Basic Operation,
Programming
" section of the manual.
8) Perform the LVDT mag and zero setup procedure. See the "Basic Operation,
Input
Setup
" section of the manual.
9) The unit is now ready for operation.
Содержание Trendsetter II
Страница 2: ...rend Setter II User s Manual T T Edmunds GAGES...
Страница 15: ...2 3 2 3 Overall E8300Unit Figure 2 1 Trendsetter II Basic Unit...
Страница 17: ...2 5 Figure 2 2 Front Panel...
Страница 19: ...2 7 Figure 2 3 Rear Panel...
Страница 24: ...2 12 Figure 2 7 E8302 2 Channel LVDT Signal Conditioning Module...
Страница 26: ...2 14 Figure 2 8 E8303 1 Channel AE Signal Conditioning Module...
Страница 30: ...3 3 Front Mounting Foot Rear Mounting Bracket Figure 3 1 Base Feet Figure 3 2 Filter Regulator Mounting...
Страница 31: ...3 4 Figure 3 3 Air Connections For Air Gaging Application Only...
Страница 53: ...4 7 Figure 4 5 A E Block...
Страница 55: ...4 9 Figure 4 6a LVDT Module E8302 current version...
Страница 57: ...4 11 Input A Polarity Jumper Input B Polarity Jumper Figure 4 7a Polarity Jumpers current version...
Страница 59: ...4 13 Output Jumpers Input Jumpers Input Type Jumpers Figure 4 8 Input Output Jumpers...
Страница 65: ...4 19 J10 J3 J4 Figure 4 10a Attenuation Jumpers current revision...