4-10
Polarity
The polarity of the input determines whether the input reads positive or negative as the
probe tip is depressed. The bargraph will rise when the tip of the probe is pressed toward
the body of the probe in the (+) setting. The polarity of input A and input B can be set
independently. For example, if a gaging fixture uses two LVDTs to measure an outside
diameter, then the polarity of both probes should be set to ("+") so that for larger part
outside diameters the readout would show a larger size.
To set input polarity position the input polarity switches on the circuit board to ("+") or
("- ") as shown in figure 4.7 or set jumpers J12 and J13 as shown in figure 4.7a. Figure
4.7a shows the current version of the jumper settings, while figure 4.7 shows the obsolete
original version.
Input A Polarity Switch
Input B Polarity Switch
Figure 4.7 - Polarity Switch, original version
Содержание Trendsetter II
Страница 2: ...rend Setter II User s Manual T T Edmunds GAGES...
Страница 15: ...2 3 2 3 Overall E8300Unit Figure 2 1 Trendsetter II Basic Unit...
Страница 17: ...2 5 Figure 2 2 Front Panel...
Страница 19: ...2 7 Figure 2 3 Rear Panel...
Страница 24: ...2 12 Figure 2 7 E8302 2 Channel LVDT Signal Conditioning Module...
Страница 26: ...2 14 Figure 2 8 E8303 1 Channel AE Signal Conditioning Module...
Страница 30: ...3 3 Front Mounting Foot Rear Mounting Bracket Figure 3 1 Base Feet Figure 3 2 Filter Regulator Mounting...
Страница 31: ...3 4 Figure 3 3 Air Connections For Air Gaging Application Only...
Страница 53: ...4 7 Figure 4 5 A E Block...
Страница 55: ...4 9 Figure 4 6a LVDT Module E8302 current version...
Страница 57: ...4 11 Input A Polarity Jumper Input B Polarity Jumper Figure 4 7a Polarity Jumpers current version...
Страница 59: ...4 13 Output Jumpers Input Jumpers Input Type Jumpers Figure 4 8 Input Output Jumpers...
Страница 65: ...4 19 J10 J3 J4 Figure 4 10a Attenuation Jumpers current revision...