34
distilled water, and wipe the mirror and surrounding areas again.
4. Dry these areas thoroughly with a clean, dry cotton swab.
5. Replace the sensor cover and the sample chamber cover.
6.
Press the “HEAT” key to return to normal operation.
7. P
ress the “MABC” key to balance the optics and return to normal operation.
A CHECK SENSOR error flag on the Display shows that the dew point sensor
control loop is not able to develop a dew layer on the mirror. The only service that
can be performed in the field is to lift the Panel and check for a loose connector in
the mirror control system, or call the Factory.
14.1.3 DRYER TUBE MAINTENANCE
When the Drier-ite
™ pellets in the Dryer Tube have changed color from blue to pink,
the drying capability has been exhausted. The dryer cylinder can then be removed
by loosening the two Velcro
™ straps. Lift the tube out and carefully disconnect the
two clear plastic tubes from the fittings at the bottom of the dryer tube.
Figure 14-2a. Open the straps Figure 14-2b. Lift out the Dryer
Figure 14-2c. Disconnect clear tubing Figure 14-2d. Replace the Dryer