20 |
P a g e
Note
If non-magnetic paint or coating is present, there is no need to remove it (unless it is thicker
than 3 mm (0.118 in). The Sharck calibration process takes that lift-off into account and
manages accordingly.
11.
In the
Sharck Assisted Calibration
dialog box, tap
Calibrate (
)
..
When the necessary data is acquired, a green check mark appears next to
Calibrate
.
Figure 28 Sharck Assisted Calibration: Calibration complete
Note
If, for any reason, you must normalize the probe again, tap
Reset
before you perform the above
procedure.
12.
Tap the red “x” on the top right corner of the dialog box to close it.