Nulling the Sharck Probe
1.
Position the Sharck probe on the plastic portion of Eddyfi’s normalization plate (REFPL-CAL-
SHARCK-AAL-LG).
Note
The normalization plate should be on a steady, flat surface and shall be at least 10 cm (4 in)
from any ferritic surface.
2.
Apply even pressure on the probe.
Figure 26 Nulling the Sharck probe
Note
To balance the probe, we strongly recommend that you press the probe onto a non-metallic
surface, instead of keeping it in the air. This avoids having one or only a few fingers compressed
at a lower position than neighboring fingers, which can affect the probe’s balance.
3.
On Reddy’s keypad, press the
Probe nulling
button
).
Alternately, in the front stage view, on the
Home
or
Calibration
ribbon, tap
Null
( ).