| 49
Maintenance and Troubleshooting
Maintaining Reddy
®
Because of its design, Reddy only requires minimal maintenance. Since Reddy has no moving parts, it
also does not require any preventive maintenance on your part. We recommend a regular inspection of
the instrument to ensure that it is properly grounded. We also strongly recommends an annual calibration
and a factory-performed preventive maintenance by an officially qualified Eddyfi technician.
Cleaning Reddy
1.
Make sure that the instrument is off and that the power cord is disconnected.
2.
To bring the instrument back to its original finish, clean it with a soft cloth.
Warning
Do not
spray the instrument with chemical cleansers or water. Doing so may lead to short circuits and
damage to the instrument.
Important
To remove stubborn stains, use a cloth moistened with soft, soapy solution. Do not use abrasives or
strong solvents as they could damage the finish. Wait until the instrument is completely dry before
connecting the power cord or cables.
Updating and Upgrading Magnifi
®
GO
Before you can perform any maintenance on Magnifi GO, you must first meet the following requirements:
•
USB mass storage device with a minimum of 4 GB free space ×1
•
Hardwired Internet connection
There are two methods of updating or upgrading Magnifi GO.
Standard Method
1.
Connect Reddy to a power outlet with the power cable.
2.
Turn on Reddy and wait for Magnifi GO to start.
3.
Download the
*.ReddyOS
file from the Eddyfi Web site.
Save the file in an easy-to-remember location on your computer.
4.
Copy the
*.ReddyOS
to the root of a USB mass storage device.
5.
Once copied, remove the mass storage device and connect it to one of Reddy’s two USB ports
A dialog box appears prompting you to proceed.
Important
Do not connect your mass storage device to the
Q
uiCk
C
opy
uSB
port.
6.
Tap
Yes
.
7.
In the list that appears, tap
OS Update
, and then tap
Update
.
The instrument restarts. A window opens prompting you to confirm whether you want to update
the system.
8.
To select
Yes
, press the keypad’s up arrow (see page 3). Press any other button to select
No
.
You are prompted to confirm again.
9.
Pres the keypad’s up arrow again.
The update process starts. This normally takes between 5 and 10 minute, depending on the speed
of your mass storage device. When the process is complete, the system restarts.
10.
Activate Windows.
See Activating Windows on page 50 for details.
Содержание Reddy
Страница 1: ...User s Guide...
Страница 9: ...ix...
Страница 10: ...x www eddyfi com General Precautions and Conventions General Precautions and Conventions...
Страница 15: ...xv General Precautions and Conventions...
Страница 16: ...2 www eddyfi com Reddy Overview Chapter 1 Reddy Overview...
Страница 27: ...13 Reddy Overview...
Страница 28: ...14 www eddyfi com MagnifiGO Overview Chapter 2 MagnifiGO Overview...
Страница 34: ...20 www eddyfi com MagnifiGO Overview Figure 2 6 Acquisition Summary Figure 2 7 Report Summary...
Страница 35: ...21 MagnifiGO Overview Figure 2 8 Save Figure 2 9 Material Database...
Страница 36: ...22 www eddyfi com Preferences Chapter 3 Preferences...
Страница 38: ...24 www eddyfi com Preferences Figure 3 3 Analysis preferences Figure 3 4 Help...
Страница 39: ...25 Preferences...
Страница 40: ...26 www eddyfi com Keypad and Keyboard Functions Chapter 4 Keypad and Keyboard Functions...
Страница 42: ...28 www eddyfi com Using Reddy with Sharck Probes Chapter 5 Using Reddy with Sharck Probes...
Страница 62: ...48 www eddyfi com Maintenance and Troubleshooting Chapter 6 Maintenance and Troubleshooting...
Страница 66: ...52 www eddyfi com Specifications Chapter 7 Specifications...
Страница 69: ...55 Specifications...
Страница 70: ...56 www eddyfi com Connector Reference Appendix A Connector Reference...
Страница 74: ...60 www eddyfi com Using the Optional Harness Appendix B Using the Optional Harness...
Страница 82: ...www eddyfi com...