38 | www.eddyfi.com
Using Reddy with Sharck Probes
Nulling the Sharck Probe
1.
Position the Sharck probe on the plastic portion of the normalization plate.
Note
The normalization plate should be on a steady, flat surface and shall be at least 10 cm (4 in) from any
ferritic surfaces.
2.
Apply even pressure on the probe.
Figure 5–22
Nulling the Sharck probe
Note
To balance the probe, we strongly recommend that you press the probe onto a non-metallic surface,
instead of keeping it in the air. This avoids having one or only a few fingers compressed at a lower
position than neighboring fingers, which can slightly affect the probe’s balance.
3.
On Reddy’s keypad, press the Probe nulling button (see page 3).
Alternatively, in the front stage view, on the
Home
or
Calibration
ribbon, tap
Null
.
Calibrating the Sharck Probe
1.
In the front stage view, on the
Calibration
ribbon, tap
Assisted Sharck
.
Figure 5–23
Sharck Assisted Calibration
2.
Position the Sharck probe on the plastic portion of the normalization plate (REFPL-CAL-SHARCK-
AAL-LG).
Note
The normalization plate should be on a steady, flat surface and shall be at least 10 cm (4 in) from any
ferritic surface.
3.
Apply even pressure on the probe.
4.
In the
Sharck Assisted Calibration
dialog box, tap
Air reference block
.
When the necessary data is acquired, a green check mark appears next to
Air reference block
.
5.
Position the Sharck probe on the aluminum portion of the normalization plate.
6.
Apply even pressure on the probe.
Содержание Reddy
Страница 1: ...User s Guide...
Страница 9: ...ix...
Страница 10: ...x www eddyfi com General Precautions and Conventions General Precautions and Conventions...
Страница 15: ...xv General Precautions and Conventions...
Страница 16: ...2 www eddyfi com Reddy Overview Chapter 1 Reddy Overview...
Страница 27: ...13 Reddy Overview...
Страница 28: ...14 www eddyfi com MagnifiGO Overview Chapter 2 MagnifiGO Overview...
Страница 34: ...20 www eddyfi com MagnifiGO Overview Figure 2 6 Acquisition Summary Figure 2 7 Report Summary...
Страница 35: ...21 MagnifiGO Overview Figure 2 8 Save Figure 2 9 Material Database...
Страница 36: ...22 www eddyfi com Preferences Chapter 3 Preferences...
Страница 38: ...24 www eddyfi com Preferences Figure 3 3 Analysis preferences Figure 3 4 Help...
Страница 39: ...25 Preferences...
Страница 40: ...26 www eddyfi com Keypad and Keyboard Functions Chapter 4 Keypad and Keyboard Functions...
Страница 42: ...28 www eddyfi com Using Reddy with Sharck Probes Chapter 5 Using Reddy with Sharck Probes...
Страница 62: ...48 www eddyfi com Maintenance and Troubleshooting Chapter 6 Maintenance and Troubleshooting...
Страница 66: ...52 www eddyfi com Specifications Chapter 7 Specifications...
Страница 69: ...55 Specifications...
Страница 70: ...56 www eddyfi com Connector Reference Appendix A Connector Reference...
Страница 74: ...60 www eddyfi com Using the Optional Harness Appendix B Using the Optional Harness...
Страница 82: ...www eddyfi com...