Chapter 2 — The pH/mV Amp
17
you want to use.) The units conversion will only apply to subsequently
recorded signals, so it is more limited than choosing units conversion on
individual blocks or pages of data. For more information about units
conversion, see the
Chart Software Manual
or
Scope Software Manual,
which are both located on the Installer CD
.
The Temperature Probe
The pH/mV Amp is supplied with a RTD type temperature probe with
100 Ω nominal resistance, to which the Temperature Amp is factory set
to produce 0 mV at 0°C and an output of 50 mV/°C. Only use these
values to calibrate the probe supplied with the pH/mV Amp. You can
substitute other RTD probes for the one provided, however the factory
calibration will no longer be exact and you will have to use two
precision temperature sources to calibrate the probe.