16
pH/mV Amp
produce 0 mV at 0°C with an output of 50 mV/°C with the RTD
temperature probe provided with the unit. If temperature accuracy is of
critical importance you should use two precision temperature sources to
calibrate the temperature probe more accurately — refer to the
Chart
or
Scope Software Manual
.
When using the pH extension with automatic temperature
compensation, the Temperature Amp signal must first be calibrated in
degrees Celsius (°C), degrees Fahrenheit (°F), or Kelvin (K).
Clicking the
button transfers the signal in the data display
area of the dialog box to the data display area of the Units Conversion
dialog box. (Use the Pause button to capture any specific signal that
Pause/Resume
button
Select Temperature from
the channel menu
Select suitable range:
5 V up to 100 °C
2 V up to 40 °C
Access Units
Conversion for
Preview of
temperature signal
0 V = 0 °C and
5 V = 100 °C
are the factory
default settings for
the RTD probe
Select
temperature
units
Figure 2–10
The temperature dialog
box (Chart software on a
Macintosh Computer)