Patient Monitor User Manual ECG/ RESP Monitoring
- 85 -
Figure 12-8 Other Setup menu
Users can access the following functions:
SMART LEAD OFF
: in
5 LEADS
mode, if the
CH1
and
CH2
can not measure
because of the lead off or other reasons, it can shift to other
LEADS
to collect a ECG
waveform.
BEAT VOL
Six selections are available:
0, 1, 2, 3, 4, 5
. “
5
” indicates maximum volume. “
0
”
indicates no sound.
PACE
ON
detected signal will be marked by a "
|
" above the ECG waveform
OFF
for non-paced patient.
NOTE:
When monitoring a patient with a pacemaker, set “
PACE
” to
ON
. If monitoring a patient
without a pacemaker, set “
PACE
” to
OFF
.
If “
PACE
” is
ON
, the system will not perform some types of ARR analysis. For detailed
information, please refer to the
Section ARR ALARM
.
CASCADE
: turn on or off
CASCADE
display. When it is set to
ON
, the ECG
waveform is display in 2 channels. This function is available only for the
NORMAL
DISPLAY
in
ECG DISPLAY
.
ECG CAL
Pick this item to start ECG calibrating process. Picking this item again can finish calibrating
process.
Users can turn on or off the Power frequency filter of 50Hz or 60Hz in
DIAGNOSTIC
mode. It can use standardized voltage to set the display width for 1mV signal. For example
X1 is for 10mm, X2 is for 20mm.
ADJUST WAVE POS