Patient Monitor User Manual Specifications
- 150 -
Appendix
I
Specifications
A1.1 Classification
Anti-electroshock Type
Class I equipment and internal powered equipment
Anti-electroshock Degree
ECG
(RESP), TEMP, IBP CF
SpO
2
, NIBP, CO
2
BF
Ingress Protection
IPX1
Disinfection/Sterilizing method
Refer to
Chapter 12 ~ Chapter 17
for details.
Working System
Continuous operation equipment
Compliant with Standards
IEC 60601-1: 1988+A1: 1991+A2: 1995; EN 60601-1:
1990+A1: 1993+A2: 1995; IEC 60601-1-2: 2001+A1:
2004; EN 60601-1-2: 2001+A1: 2006; IEC/EN
60601-2-27; IEC/EN 60601-2-30; IEC/EN 60601-2-34;
IEC/EN 60601-2-49; ISO 9919; ISO 21647; EN
12470-4; EN 1060-1; EN 1060-3; EN 1060-4;
ANSI/AAMI EC13; ANSI/AAMI SP10
A1.2 Specifications
A1.2.1 Size and Weight
Weight
< 5 kg (not including the battery and record)
A1.2.2 Environment
The monitor may not meet the performance specifications given here if stored or used outside the
specified temperature and humidity ranges.
When the monitor and related products have differing environmental specifications, the effective
range for the combined products is that range which is common to the specifications for all
products.
Temperature
Working
+5
°
C to +40
°
C
Transport and Storage
-20
°
C to +55
°
C
Humidity