Patient Monitor User Manual ECG/ RESP Monitoring
- 82 -
channel's right side. The height of 1mV bar is directly proportional to the waveform
amplitude.
NOTE:
When the input signals are too strong, the peak of the waveform may not be able to be
displayed. In this case the user may manually change the setup method of ECG
waveform according to the actual waveform so as to avoid the occurrence of the
unfavorable phenomena.
[3]. Filter method: used for displaying clearer and more detailed waveforms
There are three filter modes for selection:
DIAGNOSTIC, MONITOR
and
SURGERY
modes.
SURGERY
mode may reduce perturbance and interference from Electrosurgery
equipment. The filter method is the item applicable for both channels, which is always
displayed at the waveform place of the channel 1 ECG waveform.
NOTE:
Only in Diagnosis mode, the system can provide non-processed real signals. In Monitor
or Sugery mode, ECG waveforms may be distorted to different extents. In either of the
latter two modes, the system can only show the basic ECG and the results of ST analysis
may also be greatly affected. In Surgery mode, results of ARR analysis may be
somewhat affected. Therefore, it is suggested that in the environment where relatively
small interference exists, you’d better monitor a patient in Diagnosis mode.
[4]. Leads of channel 2: refer to [1] for detailed information.
[5]. Waveform gain of channel 2: refer to [2] for detailed information.
NOTE:
Pacemaker signal detection is marked by a "
|
" above the ECG waveform.
12.5 ECG Menu
12.5.1 ECG SETUP
Pick the ECG hot key on the screen, and the following menu will pop up.