Patient Monitor User Manual Specifications
- 156 -
Response time of Heart Rate
Meter to Change in HR
HR range: 80 bpm ~ 120 bpm
Range : 7s ~ 8s, average is 7.5s
HR range: 80bpm ~ 40bpm
Range : 7s ~ 8s, average is 7.5s
Tall
T-wave
Rejection
Exceeds ANSI/AAMI EC13-2002 Sect. 4.1.2.1 C)
minimum recommended 1.2mV T-Wave amplitude
Accuracy of Heart Rate Meter
and Response to Irregular
Rhythm
Complies with ANSI/AAMI EC13-2002 Sect.4.1.2.1 e)
The HR value displays after a stable period of 20s:
Ventricular bigeminy: 80bpm±1bpm
Slow alternating ventricular bigeminy: 60bpm±1bpm
Rapid alternating ventricular bigeminy: 120bpm±1bpm
Bidirectional systoles: 91bpm±1bpm
Non-Paced Patient
Paced Patient
ASYSTOLE
R on T
ASYSTOLE
VFIB/VTAC PVC
TACHY
COUPLET TACHY BRADY
VT>2 BRADY
PNC
BIGEMINY MISSED
BEATS
TRIGEMINY IRR
Arrhythmia analyses
VENT VBRADY
PNP
A1.2.8 RESP
Method Trans-thoracic
impedance: R-F(RA-LL), R-L (RA-LA)
Adult: 0 to 120 rpm
Neo/Ped: 0 to 150 rpm
Resolution: 1 rpm
RR measuring range
Accuracy:
±
2 rpm
Gain selection
×
0.25,
×
0.5,
×
1,
×
2,
×
3,
×
4,
×
5
Sweep 6.25mm/s,
12.5mm/s, 25.0mm/s, 50.0mm/s