TÜV Rheinland Energie und Umwelt GmbH
Luftreinhaltung
Seite 80 von 253
Report on the performance testing of the Serinus 30 ambient air quality
monitoring system manufactured by Ecotech Pty Ltd measuring CO, Report
no.: 936/21221977/D_EN
EP_Ecotech_Serinus30_en.docx
level of 8.6 µmol/mol CO. According to VDI Guideline 4202 Sheet 1, the test of the repeata-
bility standard deviation at reference point shall be performed using the limit value.
7.4
Evaluation
Table 10 details the test results for the repeatability standard deviation.
Table 10:
Repeatability standard deviation at zero and span point
7.5
Assessment
The repeatability standard deviation at zero point is 0.02 µmol/mol for system 1 and 0.02
µmol/mol for system 2. The repeatability standard deviation at span point is 0.03 µmol/mol for
system 1 and 0.00 µmol/mol for system 2.
Does this comply with the performance criterion? yes
7.6
Detailed presentation of test results
Table 11 lists the results of the individual measurements.
Requirement
≤ 0.3
0.02
0.02
≤ 0.4
0.03
0.00
0.0801
0.0697
Repeatability standard deviation s
r,z
at zero [µmol/mol]
Repeatability standard deviation s
r,ct
at c
t
[µmol/mol]
Detection limit [µmol/mol]
Device 1
Device 2
Содержание Serinus 30
Страница 120: ...Serinus 30 Kohlenmonoxid Analysator Benutzerhandbuch Version 2 1 www ecotech com...
Страница 121: ...Serinus 30 Benutzerhandbuch 2 1 Seite 2 Diese Seite wurde absichtlich frei gelassen...
Страница 135: ...Serinus 30 Benutzerhandbuch 2 1 Seite 16 Diese Seite wurde absichtlich frei gelassen...
Страница 181: ...Serinus 30 Benutzerhandbuch 2 1 Seite 62 Diese Seite wurde absichtlich frei gelassen...
Страница 197: ...Serinus 30 Benutzerhandbuch 2 1 Seite 78 Diese Seite wurde absichtlich frei gelassen...
Страница 229: ...Serinus 30 Benutzerhandbuch 2 1 Seite 110 Diese Seite wurde absichtlich frei gelassen...
Страница 237: ...Serinus 30 Benutzerhandbuch 2 1 Seite 118 Diese Seite wurde absichtlich frei gelassen...