TÜV Rheinland Energie und Umwelt GmbH
Air Pollution Control
Report on the performance testing of the Serinus 30 ambient air quality moni-
toring system manufactured by Ecotech Pty Ltd measuring CO,
Report no.: 936/21221977/D_EN
Page 13 of 253
EP_Ecotech_Serinus30_en.docx
1.3 Summary of test results
Performance criterion
Minimum requirement
Test result
Com-
pliance
Page
4
Requirements on the instrument design
4.1
General requirements
4.1.1 Measured value display Shall be available.
The measuring system is fitted with a
Necessary maintenance of the
measuring system should be
possible without larger effort, if
possible from outside.
Maintenance can be carried out with
usual tools in a reasonable time and
If the operation or the function-
al check require particular in-
struments, they shall be con-
sidered as part of the measur-
ing system and be applied in
the corresponding sub-tests
and included in the assess-
ment.
The tested instrument does not have
an internal device for functional
The set-up times and warm-up
times shall be specified in the
instruction manuals.
Set-up times and warm-up times were
The instruction manual shall
include specifications of the
manufacturer regarding the de-
sign of the system.
The specifications of the manual with
regard to instrument design are com-
4.1.6 Unintended adjustment It shall be possible to secure
the adjustment of the measur-
ing system against illicit or un-
intended adjustment during
operation.
The measuring system itself is not
protected against the unintended or
unauthorised adjustment of instru-
ment parameters. It has to be operat-
ed in a lockable measuring container.
The output signals shall be
provided digitally and/or as an-
alogue signals.
Measured signals are provided in
analogue (0–20 mA, 2–20 mA, 4–20
mA or 0–5 V) and digital form (via
TCP/IP, RS 232, USB; Bluetooth).
Содержание Serinus 30
Страница 120: ...Serinus 30 Kohlenmonoxid Analysator Benutzerhandbuch Version 2 1 www ecotech com...
Страница 121: ...Serinus 30 Benutzerhandbuch 2 1 Seite 2 Diese Seite wurde absichtlich frei gelassen...
Страница 135: ...Serinus 30 Benutzerhandbuch 2 1 Seite 16 Diese Seite wurde absichtlich frei gelassen...
Страница 181: ...Serinus 30 Benutzerhandbuch 2 1 Seite 62 Diese Seite wurde absichtlich frei gelassen...
Страница 197: ...Serinus 30 Benutzerhandbuch 2 1 Seite 78 Diese Seite wurde absichtlich frei gelassen...
Страница 229: ...Serinus 30 Benutzerhandbuch 2 1 Seite 110 Diese Seite wurde absichtlich frei gelassen...
Страница 237: ...Serinus 30 Benutzerhandbuch 2 1 Seite 118 Diese Seite wurde absichtlich frei gelassen...