Diagnostic X-ray Unit
USER MANUAL
37
Default Mode
This Mode will initialize the User Mode and APR Mode that user saved.
1. The way to access the Default Mode is same with Calibration Mode.
2. The saved values of User Mode or APR Mode will be initialized if you press the
“APR Store Key” for 2 second after choosing "S03" or "S04" by kV control key
(UP/DOWN).
Section
Setting
Function
Description
User
Mode
-
-
-
kV mAs
"S03" Mode
initializes all the
values of User
Mode.
The initialized setting is like
below.
- Buzzer “ON”
- APR “12 Mode”
- Laser & Collimator Lamp “ON”
- Sleep Time “5min”
- Power Off Time "15min"
APR
Mode
-
-
-
kV mAs
“S04” Mode
initializes all the
values of APR
Mode.
-