3. Slide the PCB out. Install a jumper on JP4.
4.
Make sure both O-rings are on the threaded connector.
5.
Slide the PCB into the enclosure until the two tangs “click”.
6.
Put the nut on and tighten ONLY ½ turn from where it is seated. If this nut is
tightened too much, the connector will crack and the enclosure will not be sealed.
7.
Connect the module to a power supply and a PC (via a CAN communication adapter)
using the cabling shown. A sensor does not have to be connected to the module.
Note that only one module is connected and a display head is not involved.