4.7.1 IMAGE ORIENTATION
To give the monitor’s image the same orientation seen through the binocular, follow these steps:
1) Loose the 3 Allen screws in the beam splitter by using the supplied 1.5mm Allen key.
2) Rotate the video camera until the orientation of the image visualized through the binoculars is
the same as the image on the monitor.
3) Adjust the 3 Allen screws of the Beam splitter to fix the camera position.
4.8 REVISION
Once installation is over and before starting-up the Ecleris Microstar OM-100 Microscope, please
go through the following check list:
Check that the voltage to be used is the same as the one stated on the back of the light
source.
Verify that the electrical network used has its grounding connection in perfect conditions
and that it makes perfect contact with the equipment.
Check that the fuse holder has a fuse installed and that it is the correct one.
Check that the ventilation grid of the light source is completely free, without obstruction.
Ensure an adequate air flow.
Verify that there are no water or liquid sources near the equipment that may damage it,
such as gases or flammable emissions in the environment.
If before connecting the Ecleris Microstar OM-100 Microscope any anomaly is detected,
please contact the technical service department of your local distributor or the closest
ECLERIS office, prior to its use
5. OPERATION
– FUNCTIONS
5.1 MICROSCOPE OPERATION
This section thoroughly describes the equipment functioning procedure.
The equipment shall only be operated by qualified personnel.