C. pH Measurements 

To take pH measurements, the pre-selected buffer solution set must lit 
up,  indicating  the  unit  is  Single  point  or  dual-point  or  three-point 
calibrated  and  ready  for  measurements.  If  buffer  solution  set  flashes, 
perform a pH calibration before taking measurements.  

a.  Measurement  with  an  ATC/Temp  probe  in  the  pH  AUTOLOCK 

mode 

1.  Connect the pH electrode to the BNC connector and the ATC/Temp 

probe to the ATC/Temp connector of the unit. The “ATC” icon will 
light up. 

2.  Press “Mode” key until “pH” icon and “AUTO” icon light up. 

3.  Rinse the pH electrode and ATC/temp probe with distilled water and 

immerse  in  the  sample  to  be  measured.  Remove  any  air  bubbles 
trapped around the probe by shaking or stirring the probe. 

4.  Press the “Mea.” key. The “WAIT” icon will start flashing. The unit is 

waiting for a stable reading. The display will track the pH value as 
sensed by the pH electrode and the ATC/Temp probe. 

5.  When the “WAIT” icon disappears, the reading is then “LOCK” and 

will not respond to further changes from the sample. The pH value 
shown  is  the  pH  value  of  the  sample  at  the  displayed  sample 
temperature. 

[Note:  For  samples  that  are  inherently  unstable,  the  unit  will  not 
AUTOLOCK. In these cases, use the pH NON- AUTOLOCK mode for 
measurements.] 

b.  Measurement with manual temperature compensation in the pH-

AUTOLOCK mode 

1.  Connect  the  pH  electrode  to  the  BNC  connector  of  the  unit.  The 

“MAN” icon will lit up. Set unit to display the sample temperature 
by pressing the up and down keys (0.0 to 100

). 

2.  Repeat steps 2~5 of “Measurement with an ATC/Temp probe in the 

pH- AUTOLOCK mode”. 

Содержание L20

Страница 1: ...Model L20 Instruction Manual Microprocessor based pH ORP Temperature Bench Top Meter Electro Chemical Devices Inc 1500 N Kellogg Dr Anaheim CA 92807 USA Telephone 1 714 695 0051 www ecdanalytical com...

Страница 2: ...Display 3 B Keys 4 Operational Procedures 5 A Buffer Set Selection 5 B pH Calibration 5 a ATC Temp probe pH AUTOLOCK mode 5 b Manual temperature compensation pH AUTOLOCK mode 6 c ATC Temp probe pH NON...

Страница 3: ...NON AUTOLOCK mode 10 d Manual temperature compensation pH NON AUTOLOCK mode 10 D Temperature Measure 10 E mV Measurements 10 a mV AUTOLOCK mode 10 b mV NON AUTOLOCK mode 11 pH Calibration Buffer Tabl...

Страница 4: ...displays pH or mV and temperature simultaneously along with user prompts and mode indicators The unit prompts the user through calibration and measurement procedures An AUTOLOCK feature for both pH an...

Страница 5: ...he AC 3 adaptor supplied with the instrument to make sure that the AC line voltage is correct If the wrong AC adaptor is supplied notify your ECD representative immediately Installing the Batteries Th...

Страница 6: ...ors 9 LOCK This will indicate that the reading is frozen during AUTOLOCK mode 4 mV Unit and mode indicators 10 EFF This will be displayed if the user is viewing the efficiency of the electrode It is r...

Страница 7: ...event accidental erasing stored memory When the Clear key is pressed all segments of the LCD will be on After about 2 seconds the unit will enter the pH AUTO mode The AUTO and CAL will be on and one o...

Страница 8: ...light up pH icon and AUTO icon light up The buffer set icon will flash 3 Rinse the pH and ATC Temp probes in distilled water and immerse them in the first buffer solution either 7 00 or 6 86 Allow tem...

Страница 9: ...splay the new electrode efficiency b Calibration with manual temperature compensation in the pH AUTOLOCK mode 1 Turn the unit on Press Clear key for 2 seconds all LCD elements will light up The meter...

Страница 10: ...ld Stand key for 5 seconds to calibrate the unit immediately calibrates the first point the selected buffer remains lit up while the remaining two buffers start to flash The unit is ready to be sloped...

Страница 11: ...up Press Mode key until pH icon is on The buffer set icon will flash 3 Rinse the pH probes in distilled water and immerse it in the first buffer solution Adjust the temperature reading to that of the...

Страница 12: ...obe by shaking or stirring the probe 4 Press the Mea key The WAIT icon will start flashing The unit is waiting for a stable reading The display will track the pH value as sensed by the pH electrode an...

Страница 13: ...ectrode to the BNC connector of the unit The MAN icon will light up 2 Repeat steps 2 4 of Measurement with an ATC Temp probe in the pH NON AUTOLOCK mode D Temperature Measure The L20 can measure tempe...

Страница 14: ...TOLOCK In this case use the pH NON AUTOLOCK mode for measurements b Measurement in the mV NON AUTOLOCK mode 1 Connect the optional combination ORP electrode to the BNC connector of the unit 2 Press Mo...

Страница 15: ...00 10 01 0 4 01 6 98 9 46 4 01 7 11 10 32 5 4 00 6 95 9 39 4 01 7 08 10 25 10 4 00 6 92 9 33 4 00 7 06 10 18 15 4 00 6 90 9 28 4 00 7 03 10 12 20 4 00 6 88 9 23 4 00 7 01 10 06 25 4 00 6 86 9 18 4 01...

Страница 16: ...not correct pH electrode slope is off by more than 30 of ideal slope Allow sufficient time for the electrode and ATC Temp probe to stabilize re press Slope key to continue the calibration procedure Ch...

Страница 17: ...range 0 C to 60 0 C pH Electrode Offset recognition 100 mV at pH 7 00 or 108 3 91 7mV at pH6 86 pH Electrode Slope recognition 30 at pH 4 00 4 01 9 18 and 10 01 Input Impedance 1012 Temperature senso...

Страница 18: ...ment shall be limited to a refund in the purchase price or at ECD s discretion to the repair or replacement of any product that proves upon ECD s examination to be defective when returned to the facto...

Страница 19: ...16...

Страница 20: ...1500 N Kellogg Dr Anaheim CA 92807 USA Phone 1 714 695 0051 www ecdanalytical com E0219...

Отзывы: