21)
4.3.2 Control Signal Wiring
Connect wires to the control connector for remote operation and remote monitoring.
Tables 4.4 and 4.5 and Figures 4.5 and 4.6 show the pin assignments.
Table 4.4 Receptacle Specifications
(Identical for all types)
Receptacle type
57FE-40500
Receptacle manufacturer
DDK. Ltd.
Adapted plug type
57-30500
Suitable wire size
AWG #24
Applied voltage for input *1
12V DC 20 mA
Contact capacity of output *2
50V DC/Maximum. 50 mA
*1. Apply voltage through pump.
*2. Apply voltage through equipment.
1
2 6
5 0
2 5
Figure 4.5 Control Signal Receptacle
(As seen from connecting side)
CAUTION
Do not wire vacant pins.
CAUTION
Apply 12V DC power for input signals on the pump side. Do
not apply this voltage on the equipment side.
The output signals are generated from an open collector
output. Apply a voltage not exceeding 50V DC on the
equipment side.
CAUTION
Wire all signals with the correct polarity
(SIGNAL/COMMON).
CAUTION
When output signals energize an inductive load, such as a
relay, insert a diode (100V. 1A class) to absorb the back
electromotive force from surge currents.
EBARA
Corporation
Содержание A07V
Страница 9: ...vii EBARA Corporation...
Страница 10: ...viii EBARA Corporation...
Страница 11: ...ix EBARA Corporation...
Страница 14: ...xii Blank page EBARA Corporation...
Страница 21: ...7 EBARA Corporation Outline Drawing A70W...
Страница 22: ...8 Outline Drawing A30W EBARA Corporation...
Страница 23: ...9 EBARA Corporation Outline Drawing A25S...
Страница 24: ...10 EBARA Corporation Outline Drawing A10S...
Страница 25: ...11 EBARA Corporation Outline Drawing A07V...
Страница 60: ...46 Blank Page EBARA Corporation...
Страница 63: ...III Label Country US Label Emergency Number 716 293 3330 EBARA Corporation...
Страница 69: ...IX Blank Page EBARA Corporation...