3. Critically ill patients should not be tested with
EASYMAX
®
Voice
Blood Glucose Meter.
4. DO NOT use during xylose absorption testing. Xylose in the blood will interfere Self-Monitoring
Blood Glucose System.
5. Interfering Substances depend on the concentration. The below substances up to the test
concentration will not affect the test results.
80 mg/dL
(4.4 mmol/L)
250 mg/dL
(13.9 mmol/L)
500 mg/dL
(27.8 mmol/L)
Ascorbic Acid
4 mg/dL
(0.26 mmol/L)
10.89%
-1.76%
4.55%
Ibuprofen
50 mg/dL (2.43 mmol/L) 3.10% 2.88%
4.62%
L-Dopa
1.8 mg/dL (0.09 mmol/L) 10.59% 7.91%
4.90%
Sodium Salicylate
50 mg/dL (3.12 mmol/L) -2.59% 9.42%
-0.84%
Tetracycline 1.5
mg/dL
(0.03
mmol/L) -5.32% 3.81%
3.20%
Tolbutamide
100 mg/dL (3.7 mmol/L)
-2.60%
12.30%
0.89%
Bilirubin-unconjugated 2.4
mg/dL (0.04 mmol/L)
-2.52%
4.05%
-0.23%
Uric acid
8 mg/dL (0.48 mmol/L) 2.71% 9.55%
-1.75%
Xylose
4 mg/dL (0.27 mmol/L)
-5.12%
-1.64%
-4.44%
Glucose
Level
Bias
Concentrations
of the interference tested