Limitations
The test strips are used for venous blood or fresh capillary whole blood samples.
1. DO NOT use serum or plasma sample.
2. DO NOT use anticoagulant NaF or potassium oxalate for venous sample preparation.
3. DO NOT use neonate blood sample.
4. Extreme humidity may affect the results. A relative humidity greater than 90% may cause
incorrect results.
5. The system should be used at temperatures between 10°C and 40°C (50°F and 104°F). Outside
this range, the system may get incorrect results.
6. DO NOT reuse the test strips. The test strips are for single use only.
7. Hematocrit: The hematocrit between 20% and 60% will not affect the results. Hematocrit below
20% may cause higher results. Hematocrit above 60% may cause lower results.
8. Altitude up to 3,048 meters above sea level has no effect on readings.
Healthcare Professionals – Please note these additional Limitations
1. If the patient has the following conditions, the result may fails:
Severe dehydration
Severe hypotension (low blood pressure)
Shock
A state of hyperglycemic-hyperosmolar state (with or without ketosis)
2. Lipemic samples: Cholesterol level up to 500 mg/dL (12.92 mmol/L) and triglycerides up to 3,000
mg/dL (33.6 mmol/L) do not affect the results. Grossly lipemic patient samples have not been
tested and are not recommended for testing with
EASYMAX
®
Voice
Blood Glucose Meter.