Connecting the PROCESSOR
1C7836 – February 1996
31
Connecting the EXHAUST DUCT from the DRYER
[1] Connect a 7.62 cm (3.0 in.) EXHAUST DUCT from the PROCESSOR to the BUILDING EXHAUST DUCT with
an adjustable gap.
[2] Check the negative exhaust flow of the building. See the Site Specifications, Publication Part No. 4B5476, for
more information.
(a) Use AIR METER TL-2431, to measure the negative static pressure in the EXHAUST DUCT. Measure
30.5 cm (12 in.) from the end that will be connected to the PROCESSOR.
(b) To obtain the required negative static pressure, adjust the gap between the BUILDING EXHAUST DUCT
and the EXHAUST DUCT from the PROCESSOR.
Table 3–2
Measuring the Negative Static Pressure
[3] Attach the EXHAUST ELBOW to the PROCESSOR with the self-tapping SCREW provided in the PREPACK
POLYBAG.
[4] Connect the EXHAUST DUCT to the EXHAUST ELBOW on the PROCESSOR.
Note
The
Kodak AUXILIARY VENTILATION FAN KIT is available for use with all Kodak PROCESSORS to aid in meeting
site specifications for ventilation. The KIT is available through Service Parts Management. Order the correct KIT for
your voltagae requirements.
Diameter of the
Negative Static Pressure, Water Head
EXHAUST DUCT
MIN
MAX
76 mm (3 in.)
0.76 mm (0.03 in.)
1.02 mm (0.04 in.)
102 mm (4 in.)
0.25 mm (0.01 in.)
0.51 mm (0.02 in.)
Part No.
Description
264503
Kodak Auxiliary Ventilation Fan Kit / 110 V (operates on 110 V AC)
8B7105
Kodak Auxiliary Ventilation Fan Kit (operates on 95 to 250 V AC, 47 to 63 Hz)
30.5 cm
(12 in.)
5.08 cm
(2 in.)
maximum
H127_0190BA
H127_0190BCB
EXHAUST ELBOW
EXHAUST
DUCT
DUCT
EXHAUST
BUILDING
EXHAUST DUCT
L TUBE
RUBBER HOSE
AIR METER TL-2431
CENTER CONNECTOR