22
23
The applicator button must not be pressed until the green indicator light is illuminated. If the button is pressed too early
(before the green indicator light is illuminated), a flash may not be generated.
❙
STAGE 6
END OF THE SESSION
• Press the on/off button of the device (icon
9
).
•
The fans will continue to run for a little under 5 minutes. Do not disconnect the mains
lead while the fans are still running. Amongst other factors, this will ensure that the
optical components are correctly cooled.
• Unlike the fans in the applicator, which will shut down after a certain time, the fans
in the base unit will continue to run until the mains lead of the device is finally dis
-
connected.
• Remove your
mask
glasses and remove the
eye
mask
from your patient.
• Remove the
gel
and clean off any surplus using micellar cleansing water.
• Using the
clean
spray, clean the glass end of the applicator (optical guide). To do
this, apply the
clean
spray in advance to a lint-free cloth, then clean the optical
guide (photo
13
).
•
It is also possible to service the base in which the applicator is fitted (by rinsing with
water, and ensuring that the base is dry before the applicator is refitted). To this end,
the base, which has a smoked black finish, must be removed by holding the two fins
on either side (photos
14
&
15
).
13
14
15
TREATMENT OF MEIBOMIAN BLEPHARITIS USING
❙
STAGE 5
FLASH TREATMENT
• The
eye mask
should be correctly adjusted on the eyes of your patient, and must be
worn throughout the treatment session.
• Any beauty spots should be covered using the
patches
.
• Apply the
gel
in a generous layer (minimum 1cm thick) on the cheekbone and the tempo-
ral areas, making sure that the
gel
does not get under the
eye mask
.
• Locate your
mask
goggles and put them on. If there are other persons close to you during
the operation of
, they may wear
mask
glasses
for their comfort.
• If necessary, use the
stop
plate to mask a substancial area which is not to receive flash
treatment (tattos).
• Position the end of the optical guide on the
gel.
• The optical guide must be in full contact with the
gel
, but not with the skin.
•
Trigger a flash by pressing the button on the front of the applicator.
•
Release the button and generate a further flash, after moving the optical guide under one
eye, starting from the inner canthus of the eye, up to the temporal area.
•
Repeat the same operation until 5 flashes have been applied under each eye, as per the
drawing on your right.
IMPORTANT:
No flash treatment must be administered unless the
eye mask
protective shells are in place. The
gel
will only be applied to the area which is to receive flash treatment once the
eye mask
shells are in place on the patient.
Under no circumstances must the
gel
penetrate below the
eye mask
shells.
NOTE:
The
gel
, the self-adhesive
patches
, the guide plate
stop
, the
mask
goggles and the
eye mask
protective shells
used must be those supplied with your
. The use of any other product may result in the risk of skin burns, and will invalidate
the guarantee of your device.
• Validating the parameters selected
(screen
I
).
This screen allows the options entered by the user to be reviewed; these options are vali-
dated by clicking on the «tick» icon.
At any time, a parameter can be modified by clicking on that parameter, or by returning to
the previous menu, using the arrow at the bottom left.