Страница 1: ...User Manual...
Страница 2: ...nging and is a practical example of this fact Your patients can now enjoy the benefits of a new treatment for meibomian blepharitis dry eye syn drome using state of the art pulsed light technology wit...
Страница 3: ...ETC The manufacturer s guarantee for parts and labour will only be valid if your device is returned in its complete original packaging All items of original packaging in which your was delivered shou...
Страница 4: ......
Страница 5: ...onal user of must ensure that only persons who fulfil the criteria for use described in the present manual will receive treatment using Before using your the following information should be read caref...
Страница 6: ...lement for less than two months On an injured burned or infected skin On skin lesions of unusual appearance which will require investigation On high risk areas blemishes birthmarks warts unhealed woun...
Страница 7: ...nection Activation card slot Applicator holding Ventilation grid Visor to protect touchscreen Mobile head Release clips for the mobile head Mobile head Applicator zone Applicator zone Air vents Trigge...
Страница 8: ...use of the device 13 Pause icon for the flash treatment session pressing this icon will place in pause mode In this mode no flash can be emitted from the applicator To exit this mode press the same i...
Страница 9: ......
Страница 10: ...plicator photo 3 Push the extractor into the mobile head to its full length photo 4 Apply pressure simultaneously to the 2 retaining lugs of the mobile head and remove the mobile head photo 5 Remove t...
Страница 11: ...nce entered these codes will not be requested a second time Connect using the mains lead The internal test procedure of will commence and will last for a few seconds Upon the completion of this phase...
Страница 12: ...n a generous layer minimum 1cm thick on the cheekbone and the tempo ral areas making sure that the gel does not get under the eye mask Locate your mask goggles and put them on If there are other perso...
Страница 13: ......
Страница 14: ...tivation card and its optic cartridge must be replaced when the number of flashes reaches zero STAGE 2 COOLING OF OPTICAL PARTS During the procedure for the replacement of the optic cartridge before s...
Страница 15: ...tridges are not symmetrical so they must be inserted in the right sense and certainly not in the other Check that the cartridge is correctly engaged PRECAUTIONS throughout the procedure for the replac...
Страница 16: ...the old activation card and insert the new card supplied with the optic cartridge in the slot provided for this purpose photo I Each activation card carries the same serial number as the optic cartrid...
Страница 17: ...on icon 7 The C key can be used to correct the figure in each box SELECT LANGUAGE screen L Click on the flag which corresponds to the required language The Language change confirmed message will be d...
Страница 18: ......
Страница 19: ...o danger or entail the risk of irreparable damage to your device Once o p t i c cartridge is removed it should be handled by the plastic parts only above all the electrodes the tube or the reflector s...
Страница 20: ...sothiazolinone 0 05 0 15 Deionized water Q S 100 For external use only Not to be used on broken burned or infected skin Keep out of reach of children The tool extractor is used for the separation of t...
Страница 21: ...of your device Never apply the clean spray directly to the internal components of the applicator APPLICATOR BODY All plastic components of the applicator must be cleaned using a clean cloth which has...
Страница 22: ...ols of the optical parts of the applicator to end SECU 03 SECU 04 SECU 05 SECU 07 SECU 09 Your device is blocked Contact our technical department by filling out the contact form on www esw vision com...
Страница 23: ...pheric pressure 70 106 kPa MARQUAGE Reference to the treatment of meibomian blepharitis dry eye syndrome using device Unique serial number of your E SWIN Rue des C tes d Orval ZA de la Pr v t 78550 Ho...
Страница 24: ...considered If the measured field strength in the location in which the device is used exceed the applicable RF compliance level above the device should be observed to verify normal operation If abnor...
Страница 25: ...or the applicator as you may risk exposure to high voltages Never use cartridges and acces sories other than those designated and sold by E Swin SAS for use with ENVIRONMENTAL PROTECTION You must abi...
Страница 26: ...ns described in this manual will invalidate the guarantee The company E Swin SAS will not accept liability for any accident arising from failure to observe the instructions contained in this manual or...