![E Instruments E4500 Скачать руководство пользователя страница 74](http://html2.mh-extra.com/html/e-instruments/e4500/e4500_operating-and-maintenance-manual_3066994074.webp)
74
K000000000EJ 023826A0 100714
11.0 PRINT
11.1 Print
PARAMETER DESCRIPTION
Enables the Print Menu. Allows to print the combustion analysis data on a paper ticket which
reports the measurement values. The printed values are those shown on the display when the
menu is enabled. This menu can be used for combustion analysis, even when recalled from the
memory, for draft, smoke, ambient gas and for tightness test results.
SEE SECTION 11.2.
The user, by means of this menu, can set the test ticket printing mode:
Copies
: Allows to set the number of printed copies and layout of the ticket. Several copies of
the test ticket can be printed, choosing among different layouts according to the
information included.
Model
:
The ticket layout selection is only valid for combustion analysis and can be chosen
among Complete, Partial and Total. Tickets for draft, smoke, ambient gas
concentration and tigthness test only allow a specific layout. Layouts for combustion
analysis are specified as described in the following:
Full:
includes a header with company data as well operator data previously
programmed in the configuration menu, measurements sampled in the combustion
analysis and, when sampled, the draft, smoke and CO ambient gas values.
Partial:
only reports the combustion analysis measurement values and information,
without any header, comments or blank lines for operator comments.
Total:
is arranged with the complete.
SEE SECTION 11.3.
Paper feed:
Feeds paper in the printer; this function is most useful when replacing the paper roll
in the printer.
Print:
Prints a graphical/alphanumeric test ticket for a complete check of the printer operation.
SEE SECTION 11.4.
Selects the printer type: internal or Bluetooth.
When Bluetooth printer is selected a pairing procedure will be needed in order to match
the printer to the instrument. The pairing procedure has to be performed only once.
SEE SECTION 11.6.
It allows the user to enter, in six lines of 24 characters the name of the Company or owner of the
device or the information regarding the latter (e.g. address, telephone number), which will be
printed in the header of the analysis report.
SEE SECTION 11.5.
In this submenu the user has the possibility to view the list of measurements that the device
performs. With the interactive keys, the user can add, delete or move a selected measurement.
SEE SECTION 11.7.
07/08/14
10:00
Report
Test
Printer
Configuration
Header
Measures list
KEY
FUNCTION
Activate the context keys shown on the
display.
Returns to the previous screen.
CONTEXT KEY
FUNCTION
Selects the available parameters.
Enters in the selected parameter setting.
Selects the available parameters.
OK
◄
►
Report
Configuration
Test
Header
Pairing
Measures list
Содержание E4500