57
K000000000EJ 023826A0 100714
In the following are listed the suggested stabilization times for the ‘on site calibration’ of the sensors.
Sensor CO:
3 minutes
Sensor NO:
3 minutes
Sensor SO
2
:
10 minutes
Sensor NO
2
:
10 minutes
Sensor CxHy:
3 minutes
Sensor CO2:
3 minutes
Messages in the 'Status' line:
saving:
the instrument is saving the
performed calibration
error:
the
sensor
has
NOT
been
recalibrated for any of the following
reasons:
- The
calibration
gas
cannot
properly reach the instrument.
- Concentration for the calibration
gas has not been set in the
relevant line ‘Applied gas’.
- The user didn’t allow for the
stabilization time to properly
elapse.
- The sensor could be damaged or
exhausted and must therefore be
replaced.
→
On site calibration
Sensor CO
07/08/14
10:00
Is
uA
2.22
Measured gas
ppm
8000.0
Applied gas
ppm
8000.0
Elapsed time
00:03:00
Status
not active
Calibrate
Ia
uA
0.17
OK
OK
On site calibration
Sensor CO
07/08/14
10:00
Is
uA
2.22
Measured gas
ppm
8000.0
Applied gas
ppm
8000.0
Elapsed time
00:03:00
Status
active
Calibrate
Ia
uA
0.17
OK
OK
WARNING
At any time the user can restore the factory calibration in the instrument by setting
the 'Status' line on 'not active'.
Содержание E4500
Страница 2: ......
Страница 20: ...20 K000000000EJ 023826A0 100714 7 4 Connection diagram AAC SO01 AA SF A AAC TA03A AAC KP01 AA AL05 AA SA08...
Страница 109: ...109 K000000000EJ 023826A0 100714 12 11 Measurements External probe Not available...
Страница 138: ......
Страница 140: ......
Страница 141: ...NOTE...
Страница 142: ...NOTE...
Страница 143: ......